Real-Time Monitoring of NIH/3T3 Cell Growth and Drug Reaction Using Impedance Biosensors and Comparison with Biological Assays

Impedance biosensors are manufactured on glass slides using a semiconductor process to monitor cell growth and cell-drug reactions in real time, and the results are compared with biological assay results to confirm the validity of impedance measurement method. Approximately 10,000 cells per well were cultured for 48 h, after which 6.67 μg/mL puromycin was injected to observe apoptosis over the following 48 h. A frequency sweep from 1 kHz to 1 MHz was performed to determine the optimal frequency range, identifying 367-440 kHz as the most sensitive for detecting impedance changes. Impedance was measured every 10 min for 96 h. Capacitance gradually increased during cell proliferation, while after drug administration, a transient increase occurred within 9 h, followed by a rapid decline, indicating cell death within 24 h. The sensor utilized Electrical Cell-substrate Impedance Sensing (ECIS) to detect real-time changes in cell status without the need for staining or destruction. Comparison with conventional biological assays such as MTS and FACS confirmed that the impedance biosensor provided higher sensitivity and quantitative accuracy in monitoring both cell proliferation and apoptosis. This study demonstrates that the developed biosensor enables label-free, non-invasive, and continuous monitoring of cellular behaviors with acceptable coincidence with 3 different biological assay results. Impedance biosensor presents a promising alternative to conventional biological assays and offers potential applications in drug screening, cytotoxicity evaluation, and real-time biological monitoring.

Authors

Institutions

Publication Details

Journal
Biosensors
Published
2025-12-01
DOI
https://doi.org/10.3390/bios15120788
Primary Topic
Microfluidic and Bio-sensing Technologies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Real-Time Monitoring of NIH/3T3 Cell Growth and Drug Reaction Using Impedance Biosensors and Comparison with Biological Assays

Seok-kyu Kim, Dahyun Kang, Yeeun Kim, Moongyu Jang et al.
Biosensors
Microfluidic and Bio-sensing Technologies
article

Real-Time Monitoring of NIH/3T3 Cell Growth and Drug Reaction Using Impedance Biosensors and Comparison with Biological Assays

Seok-kyu Kim, Dahyun Kang, Yeeun Kim, Moongyu Jang, Gayoung Lee
article en

Abstract

Impedance biosensors are manufactured on glass slides using a semiconductor process to monitor cell growth and cell-drug reactions in real time, and the results are compared with biological assay results to confirm the validity of impedance measurement method. Approximately 10,000 cells per well were cultured for 48 h, after which 6.67 μg/mL puromycin was injected to observe apoptosis over the following 48 h. A frequency sweep from 1 kHz to 1 MHz was performed to determine the optimal frequency range, identifying 367-440 kHz as the most sensitive for detecting impedance changes. Impedance was measured every 10 min for 96 h. Capacitance gradually increased during cell proliferation, while after drug administration, a transient increase occurred within 9 h, followed by a rapid decline, indicating cell death within 24 h. The sensor utilized Electrical Cell-substrate Impedance Sensing (ECIS) to detect real-time changes in cell status without the need for staining or destruction. Comparison with conventional biological assays such as MTS and FACS confirmed that the impedance biosensor provided higher sensitivity and quantitative accuracy in monitoring both cell proliferation and apoptosis. This study demonstrates that the developed biosensor enables label-free, non-invasive, and continuous monitoring of cellular behaviors with acceptable coincidence with 3 different biological assay results. Impedance biosensor presents a promising alternative to conventional biological assays and offers potential applications in drug screening, cytotoxicity evaluation, and real-time biological monitoring.

BiosensorsVol. 15(12)
Hallym University (KR)
National Research Foundation, Ministry of Trade, Industry and Energy, Korea Institute for Advancement of Technology, National Research Foundation of Korea
Openalex Percentile: Top 62%
Microfluidic and Bio-sensing Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.