Evaluation of the Reliability of Calibration Constants in Heat Flux Meters Using an ISO 5660-1 Cone Heater

This study evaluates the reliability of calibration constants for heat flux meters (HFMs) using a secondary standard under controlled thermal conditions provided by an ISO 5660-1 cone heater. Six HFMs (three new and three previously used) were examined by comparing manufacturer-provided (MFG) and secondary-standard-based (SEC) calibration constants. The bias between the MFG and SEC calibration constants varied substantially depending on the manufacturer, calibration method, service history, and the coating condition of the sensing surface. When the Bland-Altman limits of agreement were defined with the new HFMs as the reference, an HFM from a different manufacturer or calibrated by a different method was found to fall outside these limits. Although the absolute accuracy of the secondary-standard HFM has not been independently validated, the approach using the cone heater is practical for field implementation in terms of equipment accessibility, cost, and operational simplicity. Recalibration against the same secondary-standard prior to use is recommended to ensure repeatability and reproducibility across experiments and institutions.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-12-05
DOI
https://doi.org/10.3390/s25247406
Primary Topic
Flow Measurement and Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of the Reliability of Calibration Constants in Heat Flux Meters Using an ISO 5660-1 Cone Heater

Cheol-Hong Hwang, Woo-Geun Kim, Sung Chan Kim
Sensors
Flow Measurement and Analysis
article

Evaluation of the Reliability of Calibration Constants in Heat Flux Meters Using an ISO 5660-1 Cone Heater

Cheol-Hong Hwang, Woo-Geun Kim, Sung Chan Kim
article en

Abstract

This study evaluates the reliability of calibration constants for heat flux meters (HFMs) using a secondary standard under controlled thermal conditions provided by an ISO 5660-1 cone heater. Six HFMs (three new and three previously used) were examined by comparing manufacturer-provided (MFG) and secondary-standard-based (SEC) calibration constants. The bias between the MFG and SEC calibration constants varied substantially depending on the manufacturer, calibration method, service history, and the coating condition of the sensing surface. When the Bland-Altman limits of agreement were defined with the new HFMs as the reference, an HFM from a different manufacturer or calibrated by a different method was found to fall outside these limits. Although the absolute accuracy of the secondary-standard HFM has not been independently validated, the approach using the cone heater is practical for field implementation in terms of equipment accessibility, cost, and operational simplicity. Recalibration against the same secondary-standard prior to use is recommended to ensure repeatability and reproducibility across experiments and institutions.

SensorsVol. 25(24)
Daejeon University (KR), Kyungil University (KR)
Ministry of Land, Infrastructure and Transport, Korea Agency for Infrastructure Technology Advancement, Korea Institute of Energy Technology Evaluation and Planning
Openalex Percentile: Top 57%
Flow Measurement and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of the Reliability of Calibration Constants in Heat Flux Meters Using an ISO 5660-1 Cone Heater — Cheol-Hong Hwang, Woo-Geun Kim, et al. · Sensors (2025) | TGRS Research Map | TGRS