On the influence of reference sample properties on magnetic force microscopy calibrations

Magnetic force microscopy (MFM) allows the characterization of magnetic stray field distributions with high sensitivity and spatial resolution. Based on a suitable calibration procedure, MFM can also yield quantitative magnetic field values. This process typically involves measuring a reference sample to determine the distribution of the tip's stray field or stray field gradient at the sample surface. This distribution is called the tip transfer function (TTF) and is derived through regularized deconvolution in Fourier space. The properties of the reference sample and the noise characteristics of the detection system significantly influence the derived TTF, thereby limiting its validity range. In a recent study, the tip stray field distribution, and hence the TTF, of an MFM tip was independently measured in real space using a nitrogen vacancy center as a quantum sensor, revealing considerable discrepancies with the reference-sample-based TTF. Here, we analyze the influence of the feature distribution of the reference sample and the MFM measurement parameters on the resulting TTF. We explain the observed differences between quantum-calibrated stray field distributions and the classical approach by attributing them to a loss of information due to missing or suppressed spectral components. Furthermore, we emphasize the importance of the spectral coverage of the TTF. Our findings indicate that for high-quality reconstruction of the stray field of a sample under test (SUT), it is more critical to ensure a strong overlap of frequency components between the reference sample and the SUT than to achieve an accurate real-space reconstruction of the tip stray field distribution.

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-12-01
DOI
https://doi.org/10.1063/5.0288740
Primary Topic
Force Microscopy Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On the influence of reference sample properties on magnetic force microscopy calibrations

S. Sievers, H. W. Schumacher, Baha Sakar, Christopher Habenschaden
Review of Scientific Instruments
Force Microscopy Techniques and Applications
article

On the influence of reference sample properties on magnetic force microscopy calibrations

S. Sievers, H. W. Schumacher, Baha Sakar, Christopher Habenschaden
article en

Abstract

Magnetic force microscopy (MFM) allows the characterization of magnetic stray field distributions with high sensitivity and spatial resolution. Based on a suitable calibration procedure, MFM can also yield quantitative magnetic field values. This process typically involves measuring a reference sample to determine the distribution of the tip's stray field or stray field gradient at the sample surface. This distribution is called the tip transfer function (TTF) and is derived through regularized deconvolution in Fourier space. The properties of the reference sample and the noise characteristics of the detection system significantly influence the derived TTF, thereby limiting its validity range. In a recent study, the tip stray field distribution, and hence the TTF, of an MFM tip was independently measured in real space using a nitrogen vacancy center as a quantum sensor, revealing considerable discrepancies with the reference-sample-based TTF. Here, we analyze the influence of the feature distribution of the reference sample and the MFM measurement parameters on the resulting TTF. We explain the observed differences between quantum-calibrated stray field distributions and the classical approach by attributing them to a loss of information due to missing or suppressed spectral components. Furthermore, we emphasize the importance of the spectral coverage of the TTF. Our findings indicate that for high-quality reconstruction of the stray field of a sample under test (SUT), it is more critical to ensure a strong overlap of frequency components between the reference sample and the SUT than to achieve an accurate real-space reconstruction of the tip stray field distribution.

Review of Scientific InstrumentsVol. 96(12)
Physikalisch-Technische Bundesanstalt (DE), Leipzig University (DE)
Openalex Percentile: Top 69%
Force Microscopy Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.