Concept of a new bolometer based on soft x-ray–UV/IR conversion with discrete lines of sight

We present the concept of a new bolometer based on soft x-ray-UV to IR conversion with a high-Z foil and individual Lines of Sight (LOS). A solid block collimator is used with drilled channels defining the LOS. Each LOS is equipped with a separate metallic circular carbonized foil of 2 μm thickness, and the edge of the foil is cooled to maintain a constant temperature. Laboratory experiments with carbonized gold foils have shown that it is possible to obtain a calibration curve for the IR response as a function of incoming UV power in stationary mode. The IR signal is found to be proportional to the incoming UV signal. Levels as low as 50 μW of UV incoming power have been detected in unfavorable conditions (in air, at room temperature) with both an IR camera and industrial low-cost thermopiles. In addition, there is no crosstalk between two adjacent detectors. An analytical model confirms that the IR power is proportional to the X-UV input power and gives the physical dependences of the proportionality factor in the stationary mode. The time-dependent solution allows us to get the radiative power as a function of the time varying voltage measured at the output of the thermopiles. Using a platinum foil and the parameters of an industrial low-cost thermopile, we find that levels as low as five μW should be measured. A prototype is under construction to be tested in the WEST tokamak. It consists of two lines of sight observing the low divertor and the center of the plasma. The interest of this system is that it can be rendered compatible with ITER or DEMO. Another point is that, due to its simplicity, the cost of the diagnostic can be reduced to a fraction of the price of standard metallic resistive bolometers used in tokamaks to measure the radiated power [Mast et al., Rev. Sci. Instrum. 62, 744 (1991)].

Authors

Institutions

Publication Details

Journal
Review of Scientific Instruments
Published
2025-12-01
DOI
https://doi.org/10.1063/5.0291613
Primary Topic
Magnetic confinement fusion research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Concept of a new bolometer based on soft x-ray–UV/IR conversion with discrete lines of sight

P. Devynck, G. Antar
Review of Scientific Instruments
Magnetic confinement fusion research
article

Concept of a new bolometer based on soft x-ray–UV/IR conversion with discrete lines of sight

P. Devynck, G. Antar
article en

Abstract

We present the concept of a new bolometer based on soft x-ray-UV to IR conversion with a high-Z foil and individual Lines of Sight (LOS). A solid block collimator is used with drilled channels defining the LOS. Each LOS is equipped with a separate metallic circular carbonized foil of 2 μm thickness, and the edge of the foil is cooled to maintain a constant temperature. Laboratory experiments with carbonized gold foils have shown that it is possible to obtain a calibration curve for the IR response as a function of incoming UV power in stationary mode. The IR signal is found to be proportional to the incoming UV signal. Levels as low as 50 μW of UV incoming power have been detected in unfavorable conditions (in air, at room temperature) with both an IR camera and industrial low-cost thermopiles. In addition, there is no crosstalk between two adjacent detectors. An analytical model confirms that the IR power is proportional to the X-UV input power and gives the physical dependences of the proportionality factor in the stationary mode. The time-dependent solution allows us to get the radiative power as a function of the time varying voltage measured at the output of the thermopiles. Using a platinum foil and the parameters of an industrial low-cost thermopile, we find that levels as low as five μW should be measured. A prototype is under construction to be tested in the WEST tokamak. It consists of two lines of sight observing the low divertor and the center of the plasma. The interest of this system is that it can be rendered compatible with ITER or DEMO. Another point is that, due to its simplicity, the cost of the diagnostic can be reduced to a fraction of the price of standard metallic resistive bolometers used in tokamaks to measure the radiated power [Mast et al., Rev. Sci. Instrum. 62, 744 (1991)].

Review of Scientific InstrumentsVol. 96(12)
Commissariat à l'Énergie Atomique et aux Énergies Alternatives (FR), CEA Cadarache (FR), American University of Beirut (LB)
Openalex Percentile: Top 83%
Magnetic confinement fusion research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.