Best Practices for High-Quality Anterior Segment Optical Coherence Tomography Imaging of Eyes with the Port Delivery Platform Implant

<b>Background/Objectives:</b> Anterior segment optical coherence tomography (AS-OCT) is a non-invasive imaging modality used to evaluate anterior segment features. This report aims to inform clinicians of best practices to obtain high-resolution AS-OCT images of anterior segment features of eyes implanted with the Port Delivery Platform (PDP). <b>Methods:</b> In PDP trials, AS-OCT imaging of the anterior segment was performed using Heidelberg Spectralis OCT (Heidelberg Engineering GmbH, Heidelberg, Germany) equipped with the Anterior Segment Module. Images from over 2500 separate study visits were obtained using standardized imaging parameters. The following stepwise approach was recommended to properly orient the volume scans over the extrascleral flange of the PDP ensuring that the scans are centered on the implant with adequate depth: (1) the volume scan was aligned such that: (i) the long axis of the scan was oriented parallel to the implant flange long axis, (ii) the imaging field was centered on the implant septum center, and (iii) it covered the entirety of the implant, with equal margins on either side of the implant flange; (2) the depth of the scan focus was adjusted to ensure that the conjunctiva and Tenon's capsule over the overmold, as well as sclera under the implant flange, and the septum of the implant, were captured; and (3) Steps 1 and 2 were then repeated after the scan orientation was changed so that the short-axis scans were oriented parallel to the implant flange short axis. <b>Results:</b> Utilization of AS-OCT during clinical development of the PDP allowed visualization of anterior segment features, including the conjunctiva, Tenon's capsule, and sclera, surrounding the PDP. Overall, these best practices enabled detailed structural imaging of the implant's interface with surrounding ocular tissues. Common errors resulting in poor AS-OCT image acquisition included off-center raster scans, scans not being aligned parallel to the long and/or short axes of the PDP implant, or not being oriented along the implant axes, and inappropriate scan depths. <b>Conclusions:</b> Application of a standardized AS-OCT imaging procedure was used to obtain high-quality, high-resolution images of anterior segment features in the presence of the PDP implant. The best practices reported are not a requirement for managing eyes with the PDP, but a recommendation for how to obtain high-quality images of the anterior segment of eyes with the PDP implant.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-12-05
DOI
https://doi.org/10.3390/diagnostics15243098
Primary Topic
Retinal and Macular Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Best Practices for High-Quality Anterior Segment Optical Coherence Tomography Imaging of Eyes with the Port Delivery Platform Implant

Matthew Ohr, Andrés Emanuelli, Paul Latkany, Alisa J. Prager et al.
Diagnostics
Retinal and Macular Surgery
article

Best Practices for High-Quality Anterior Segment Optical Coherence Tomography Imaging of Eyes with the Port Delivery Platform Implant

Matthew Ohr, Andrés Emanuelli, Paul Latkany, Alisa J. Prager, Glenn J. Jaffe, J. C. Castro, Rick Laoprasert
article en

Abstract

<b>Background/Objectives:</b> Anterior segment optical coherence tomography (AS-OCT) is a non-invasive imaging modality used to evaluate anterior segment features. This report aims to inform clinicians of best practices to obtain high-resolution AS-OCT images of anterior segment features of eyes implanted with the Port Delivery Platform (PDP). <b>Methods:</b> In PDP trials, AS-OCT imaging of the anterior segment was performed using Heidelberg Spectralis OCT (Heidelberg Engineering GmbH, Heidelberg, Germany) equipped with the Anterior Segment Module. Images from over 2500 separate study visits were obtained using standardized imaging parameters. The following stepwise approach was recommended to properly orient the volume scans over the extrascleral flange of the PDP ensuring that the scans are centered on the implant with adequate depth: (1) the volume scan was aligned such that: (i) the long axis of the scan was oriented parallel to the implant flange long axis, (ii) the imaging field was centered on the implant septum center, and (iii) it covered the entirety of the implant, with equal margins on either side of the implant flange; (2) the depth of the scan focus was adjusted to ensure that the conjunctiva and Tenon's capsule over the overmold, as well as sclera under the implant flange, and the septum of the implant, were captured; and (3) Steps 1 and 2 were then repeated after the scan orientation was changed so that the short-axis scans were oriented parallel to the implant flange short axis. <b>Results:</b> Utilization of AS-OCT during clinical development of the PDP allowed visualization of anterior segment features, including the conjunctiva, Tenon's capsule, and sclera, surrounding the PDP. Overall, these best practices enabled detailed structural imaging of the implant's interface with surrounding ocular tissues. Common errors resulting in poor AS-OCT image acquisition included off-center raster scans, scans not being aligned parallel to the long and/or short axes of the PDP implant, or not being oriented along the implant axes, and inappropriate scan depths. <b>Conclusions:</b> Application of a standardized AS-OCT imaging procedure was used to obtain high-quality, high-resolution images of anterior segment features in the presence of the PDP implant. The best practices reported are not a requirement for managing eyes with the PDP, but a recommendation for how to obtain high-quality images of the anterior segment of eyes with the PDP implant.

DiagnosticsVol. 15(24)
Duke University (US), University of Puerto Rico System (PR), The Ohio State University (US), University of Puerto Rico at Arecibo (PR)
Openalex Percentile: Top 46%
Retinal and Macular Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.