Standardized and Quantitative ICG Perfusion Assessment: Feasibility and Reproducibility in a Multicentre Setting

Indocyanine green near-infrared fluorescence (ICG-NIRF) imaging is widely used to assess tissue perfusion, yet its subjective interpretation limits correlation with postoperative parathyroid function. To address this, the Workflow model for ICG-angiography integrating Standardization and Quantification (WISQ) was developed. This exploratory prospective multicenter study evaluated the reproducibility of WISQ in adults undergoing total thyroidectomy at two Dutch university centres. Patients with contraindications to ICG or prior neck surgery were excluded. Intraoperative imaging used standardized camera settings with blood volume-adjusted ICG dosing, and perfusion curves were analyzed using predefined regions of interest. Eighty patients were included. Significant inter-centre variability was observed in maximum fluorescence intensity, inflow slope, and outflow slope (<i>n</i> = 30). At the lead centre, outflow was the most promising predictor of postoperative hypoparathyroidism (HPT) (median -0.33 [IQR -0.49--0.15] a.f.u./s for HPT vs. -0.68 [-0.91--0.41], <i>n</i> = 17, <i>p</i> = 0.08), although no parameter significantly predicted HPT. Repeated ICG injections consistently produced lower maximal intensities irrespective of injection rate, and reproducible curves were achieved only when ICG was freshly dissolved at 0.5 mg/mL instead of 2.5 mg/mL. These findings indicate that ICG concentration and injection technique influence perfusion kinetics and underscore the need to update WISQ with standardized injection dilution to improve its clinical utility.

Authors

Institutions

Publication Details

Journal
Life
Published
2025-12-05
DOI
https://doi.org/10.3390/life15121868
Primary Topic
Thyroid and Parathyroid Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Standardized and Quantitative ICG Perfusion Assessment: Feasibility and Reproducibility in a Multicentre Setting

Wiktor Szymański, Tessa M. van Ginhoven, Wendy Kelder, Schelto Kruijff et al.
Life
Thyroid and Parathyroid Surgery
article

Standardized and Quantitative ICG Perfusion Assessment: Feasibility and Reproducibility in a Multicentre Setting

Wiktor Szymański, Tessa M. van Ginhoven, Wendy Kelder, Schelto Kruijff, Sam P. J. van Dijk, Gooitzen M. van Dam, Liesbeth Jansen, Hugo M. Schouw, Eline A. Feitsma, Floris P. Tange, Jasper Vonk, Wido Heeman, Milou E. Noltes, Jan H. Koetje, Abbey Schepers, John Hoffman
article en

Abstract

Indocyanine green near-infrared fluorescence (ICG-NIRF) imaging is widely used to assess tissue perfusion, yet its subjective interpretation limits correlation with postoperative parathyroid function. To address this, the Workflow model for ICG-angiography integrating Standardization and Quantification (WISQ) was developed. This exploratory prospective multicenter study evaluated the reproducibility of WISQ in adults undergoing total thyroidectomy at two Dutch university centres. Patients with contraindications to ICG or prior neck surgery were excluded. Intraoperative imaging used standardized camera settings with blood volume-adjusted ICG dosing, and perfusion curves were analyzed using predefined regions of interest. Eighty patients were included. Significant inter-centre variability was observed in maximum fluorescence intensity, inflow slope, and outflow slope (<i>n</i> = 30). At the lead centre, outflow was the most promising predictor of postoperative hypoparathyroidism (HPT) (median -0.33 [IQR -0.49--0.15] a.f.u./s for HPT vs. -0.68 [-0.91--0.41], <i>n</i> = 17, <i>p</i> = 0.08), although no parameter significantly predicted HPT. Repeated ICG injections consistently produced lower maximal intensities irrespective of injection rate, and reproducible curves were achieved only when ICG was freshly dissolved at 0.5 mg/mL instead of 2.5 mg/mL. These findings indicate that ICG concentration and injection technique influence perfusion kinetics and underscore the need to update WISQ with standardized injection dilution to improve its clinical utility.

LifeVol. 15(12)
Leiden University (NL), University Medical Center Groningen (NL), University of Groningen (NL), Fryske Akademy (NL), Medisch Centrum Leeuwarden (NL), Karolinska Institutet (SE), Tracer Technologies (United States) (US), Martini Ziekenhuis (NL), Erasmus MC Cancer Institute (NL), Erasmus University Rotterdam (NL)
Openalex Percentile: Top 42%
Thyroid and Parathyroid Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.