Impact of Gantry, Collimator, and Couch Rotational Errors in Radiation Therapy Devices on the Irradiation Accuracy of Off-isocenter Targets and Evaluation of TG142 Tolerance

The effects of the collimator and couch were substantial, with positional deviations exceeding 1 mm even for a 0.5° rotation error. The influence of the gantry was relatively small and dependent on the target configuration. For irradiation of off-axis targets, the TG142 tolerance of ±1.0° should be regarded as a minimum standard that must be strictly observed regardless of the type of linear accelerator. However, depending on the target arrangement, clinically adequate margins may not be ensured. These findings suggest the necessity of applying stricter criteria according to target configuration and emphasize the importance of regular quality assurance.

Authors

Institutions

Publication Details

Journal
Japanese Journal of Radiological Technology
Published
2025-12-03
DOI
https://doi.org/10.6009/jjrt.26-1566
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Gantry, Collimator, and Couch Rotational Errors in Radiation Therapy Devices on the Irradiation Accuracy of Off-isocenter Targets and Evaluation of TG142 Tolerance

Takahiro Nakayama, Yoshinori Tanabe, Yasushi Fujii
Japanese Journal of Radiological Technology
Advanced Radiotherapy Techniques
article

Impact of Gantry, Collimator, and Couch Rotational Errors in Radiation Therapy Devices on the Irradiation Accuracy of Off-isocenter Targets and Evaluation of TG142 Tolerance

Takahiro Nakayama, Yoshinori Tanabe, Yasushi Fujii
article en

Abstract

The effects of the collimator and couch were substantial, with positional deviations exceeding 1 mm even for a 0.5° rotation error. The influence of the gantry was relatively small and dependent on the target configuration. For irradiation of off-axis targets, the TG142 tolerance of ±1.0° should be regarded as a minimum standard that must be strictly observed regardless of the type of linear accelerator. However, depending on the target arrangement, clinically adequate margins may not be ensured. These findings suggest the necessity of applying stricter criteria according to target configuration and emphasize the importance of regular quality assurance.

Japanese Journal of Radiological TechnologyVol. 82(2)
Okayama University (JP), Chugoku Central Hospital (JP)
Openalex Percentile: Top 53%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Impact of Gantry, Collimator, and Couch Rotational Errors in Radiation Therapy Devices on the Irradiation Accuracy of Off-isocenter Targets and Evaluation of TG142 Tolerance — Takahiro Nakayama, Yoshinori Tanabe, et al. · Japanese Journal of Radiological Technology (2025) | TGRS Research Map | TGRS