Deep Learning HASTE for Upper Abdominal MRI: Improved Image Quality, Speed, and Energy Efficiency in a Prospective Study

This prospective study aimed to perform a qualitative and quantitative comparison of deep learning (DL) and conventional T2-weighted Half-Fourier Acquisition Single-shot Turbo spin Echo (HASTE) sequences for 3T MRI acquisition of the upper abdomen. From January 2024 to April 2024, 166 patients (60 ± 14 years) scheduled for MRI of the upper abdomen were prospectively enrolled. Each patient underwent two MRI examinations: one using a conventional T2-weighted HASTE sequence, followed by a fast T2-weighted HASTE sequence reconstructed with DL. Image quality, anatomical structure visualization, and diagnostic performance were independently assessed by three readers using a 5-point Likert scale. Quantitative analysis included measurements of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) for both sequences. Additionally, radiomic features were extracted and analyzed for significant variations. Interreader agreement was evaluated using Fleiss' Kappa. The DL HASTE sequence showed significantly superior overall image quality (p < 0.001), fewer artifacts (p < 0.001), and improved delineation of anatomical structures (p < 0.01) compared to the conventional T2-weighted HASTE sequence. DL sequences exhibited better SNR (p < 0.001), whereas CNR values did not show a difference between the two acquisition types. Radiomics feature analysis unveiled significant differences in contrast and gray-level characteristics (p ≤ 0.001). DL HASTE demonstrated a significant time reduction of 62.5% together with significant energy cost savings of 0.34 kW per scan compared to the conventional sequence acquisition. The DL HASTE sequence enhanced image quality and diagnostic confidence while minimizing artifacts, time, and energy costs, enabling a more accurate detection of pathologies than the conventional T2-weighted product solution with potential clinical impact.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-12-04
DOI
https://doi.org/10.1002/nbm.70183
Primary Topic
Radiomics and Machine Learning in Medical Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Deep Learning HASTE for Upper Abdominal MRI: Improved Image Quality, Speed, and Energy Efficiency in a Prospective Study

Scherwin Mahmoudi, Daniel Dahm, Christian Booz, Salah Al-Saleh et al.
NMR in Biomedicine
Radiomics and Machine Learning in Medical Imaging
article

Deep Learning HASTE for Upper Abdominal MRI: Improved Image Quality, Speed, and Energy Efficiency in a Prospective Study

Scherwin Mahmoudi, Daniel Dahm, Christian Booz, Salah Al-Saleh, Tatjana Gruber‐Rouh, Christophe Weber, Tommaso D’Angelo, Vitali Koch, P Konrad, Thomas Joseph Vogl, Dominik M. Nickel, Stefan Zeuzem, Ralph Strecker, Leon David Gruenewald, Renate Hammerstingl, Armin Wiegering, Jennifer Gotta, Simon Bernatz, Elena Höhne, Katrin Eichler, Simon S. Martin, Philipp Reschke, Ralph Sinkus, Sebastian M Haberkorn, Ahmed Ait Bachir
article en

Abstract

This prospective study aimed to perform a qualitative and quantitative comparison of deep learning (DL) and conventional T2-weighted Half-Fourier Acquisition Single-shot Turbo spin Echo (HASTE) sequences for 3T MRI acquisition of the upper abdomen. From January 2024 to April 2024, 166 patients (60 ± 14 years) scheduled for MRI of the upper abdomen were prospectively enrolled. Each patient underwent two MRI examinations: one using a conventional T2-weighted HASTE sequence, followed by a fast T2-weighted HASTE sequence reconstructed with DL. Image quality, anatomical structure visualization, and diagnostic performance were independently assessed by three readers using a 5-point Likert scale. Quantitative analysis included measurements of signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) for both sequences. Additionally, radiomic features were extracted and analyzed for significant variations. Interreader agreement was evaluated using Fleiss' Kappa. The DL HASTE sequence showed significantly superior overall image quality (p < 0.001), fewer artifacts (p < 0.001), and improved delineation of anatomical structures (p < 0.01) compared to the conventional T2-weighted HASTE sequence. DL sequences exhibited better SNR (p < 0.001), whereas CNR values did not show a difference between the two acquisition types. Radiomics feature analysis unveiled significant differences in contrast and gray-level characteristics (p ≤ 0.001). DL HASTE demonstrated a significant time reduction of 62.5% together with significant energy cost savings of 0.34 kW per scan compared to the conventional sequence acquisition. The DL HASTE sequence enhanced image quality and diagnostic confidence while minimizing artifacts, time, and energy costs, enabling a more accurate detection of pathologies than the conventional T2-weighted product solution with potential clinical impact.

NMR in BiomedicineVol. 39(1)
University of Messina (IT), Goethe University Frankfurt (DE), Siemens (Germany) (DE), Inserm (FR), Université Paris Cité (FR), Heidelberg University (DE), University Hospital Heidelberg (DE), University Hospital Frankfurt (DE), Laboratoire de Recherche Vasculaire Translationnelle (FR), Siemens Healthineers (Germany) (DE)
Siemens Healthineers
Openalex Percentile: Top 46%
Radiomics and Machine Learning in Medical Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.