Effects of subject motion and acquisition time on cone‐beam CT‐guided adaptive therapy for novel 6 second fast scanning protocols
CB-16.6 and CB-6 provided high-quality images for ART. Compared to FB-0.5, fast CBCT exhibited increased blurring and streaking but significantly less geometric distortion with increasing motion, leading to reduced dose delivery error in a simulated ART workflow. The complex 3D motion phantom study supported trends observed in the 1D translational phantom study. Faster CBCT (6 vs. 16.6 s) reduced blur, improving auto-contouring performance for both periodic and non-periodic motions, but the interplay between the motion period and 6 s image acquisition must be carefully managed.