Determination of Rare Earth Elements in Soil and Sediments by Microwave Digestion-inductively Coupled Plasma Mass Spectrometry with HNO<sub>3</sub>-HF

As strategic national mineral resources, rare earth elements are prioritized in the national soil background survey program. Consequently, the development of analytical methods enabling rapid and accurate determination of rare earth elements in soils and sediments is of critical importance. A method for the determination of rare earth elements in soil and sediments by microwave digestion-inductively coupled plasma mass spectrometry with HNO<sub>3</sub>-HF was proposed. The samples were completed decomposed by a mixture acid of 3 mL HNO<sub>3</sub>- 2 mL HF in microwave digestion tubes for 0.1000 g of sample. The microwave digestion instrument is programmed to rise temperature (5 min to 120 ℃, keep for 5 min; 5 min to 150 ℃, keep for 5 min; 5 min to 190 ℃, keep for 30 min). The optimization experiment results showed that the best amount of hydrofluoric acid was 2 mL. The removal of HF was required, with Ge、Rh、Re as the mixed internal standard elements, using He collision mode to eliminate the interference of polyatomic ions, the sample solution was directly determined by inductively coupled plasma mass spectrometry which was equipped with HF-resistant injection system. The correlation coefficients of the calibration curves for all 16 rare earth elements were greater than 0.999 7, with detection limits ranging from 0.004 μg/g to 0.04 μg/g. The rare earth elements in the certified reference materials were analyzed in accordance with the established experimental methodology. The measured values fell within the acceptable error range of the certified values, with relative standard deviations (<italic>RSD</italic>, <italic>n</italic>=5) ranging from 0.62% to 4.41%. The logarithmic concentration difference (Δlg<italic>C</italic>) remained consistently below 0.06, confirming the accuracy and reliability of the analytical method.

Authors

Publication Details

Journal
Chinese Rare Earths
Published
2025-12-01
DOI
https://doi.org/10.3724/s1004-0277.202506012
Primary Topic
Extraction and Separation Processes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determination of Rare Earth Elements in Soil and Sediments by Microwave Digestion-inductively Coupled Plasma Mass Spectrometry with HNO3-HF

Jianwei Meng, Jin Zou, L. X. Li, Ning Zhang et al.
Chinese Rare Earths
Extraction and Separation Processes
article

Determination of Rare Earth Elements in Soil and Sediments by Microwave Digestion-inductively Coupled Plasma Mass Spectrometry with HNO3-HF

Jianwei Meng, Jin Zou, L. X. Li, Ning Zhang, A Liu, Lu Liu, Min Ge
article en

Abstract

As strategic national mineral resources, rare earth elements are prioritized in the national soil background survey program. Consequently, the development of analytical methods enabling rapid and accurate determination of rare earth elements in soils and sediments is of critical importance. A method for the determination of rare earth elements in soil and sediments by microwave digestion-inductively coupled plasma mass spectrometry with HNO<sub>3</sub>-HF was proposed. The samples were completed decomposed by a mixture acid of 3 mL HNO<sub>3</sub>- 2 mL HF in microwave digestion tubes for 0.1000 g of sample. The microwave digestion instrument is programmed to rise temperature (5 min to 120 ℃, keep for 5 min; 5 min to 150 ℃, keep for 5 min; 5 min to 190 ℃, keep for 30 min). The optimization experiment results showed that the best amount of hydrofluoric acid was 2 mL. The removal of HF was required, with Ge、Rh、Re as the mixed internal standard elements, using He collision mode to eliminate the interference of polyatomic ions, the sample solution was directly determined by inductively coupled plasma mass spectrometry which was equipped with HF-resistant injection system. The correlation coefficients of the calibration curves for all 16 rare earth elements were greater than 0.999 7, with detection limits ranging from 0.004 μg/g to 0.04 μg/g. The rare earth elements in the certified reference materials were analyzed in accordance with the established experimental methodology. The measured values fell within the acceptable error range of the certified values, with relative standard deviations (<italic>RSD</italic>, <italic>n</italic>=5) ranging from 0.62% to 4.41%. The logarithmic concentration difference (Δlg<italic>C</italic>) remained consistently below 0.06, confirming the accuracy and reliability of the analytical method.

Chinese Rare EarthsVol. 46(6)
Openalex Percentile: Top 60%
Extraction and Separation Processes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.