A novel pupillometer smartphone application demonstrates accurate results compared with a validated clinical pupillometer: The Pupillary Light Response Validation (PLR-V) Study

The pupillary light response is a sensitive biomarker commonly used for neuromonitoring in intensive care units worldwide. Automated pupillometers have demonstrated increased inter-rater reliability compared with manual assessments by clinicians, allowing for objective repeated assessment. The purpose of this study was directly to compare the raw values and reproducibility of results from a novel automated pupillometer smartphone application, MindMirror, against an existing validated pupillometer, the Neuroptics NPi-200. This study was a prospective cohort study comparing pupillary light response values from three consecutive readings from the MindMirror application and the Neuroptics NPi-200 pupillometer in 66 well subjects. Initial diameter values for NPi-200 and MindMirror did not demonstrate a clinically significant mean error (−0.37 mm ± 0.51 mm) with comparable inter-sample repeatability. Other parameters suggestive of reduced constriction amplitude and velocity, were likely to be related to the clinical conditions of collection. The data suggested that the novel MindMirror smartphone application provided accurate results that were predictable and comparable to the gold standard device for automated pupillometry, and may therefore have utility as a portable accessible device for the rapid assessment of pupillary light response, particularly outside the intensive care unit.

Authors

Institutions

Publication Details

Journal
Anaesthesia and Intensive Care
Published
2025-12-04
DOI
https://doi.org/10.1177/0310057x251381378
Primary Topic
Intraoperative Neuromonitoring and Anesthetic Effects
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel pupillometer smartphone application demonstrates accurate results compared with a validated clinical pupillometer: The Pupillary Light Response Validation (PLR-V) Study

William E. Davies, Paul M Middleton, J. Schmidt
Anaesthesia and Intensive Care
Intraoperative Neuromonitoring and Anesthetic Effects
article

A novel pupillometer smartphone application demonstrates accurate results compared with a validated clinical pupillometer: The Pupillary Light Response Validation (PLR-V) Study

William E. Davies, Paul M Middleton, J. Schmidt
article en

Abstract

The pupillary light response is a sensitive biomarker commonly used for neuromonitoring in intensive care units worldwide. Automated pupillometers have demonstrated increased inter-rater reliability compared with manual assessments by clinicians, allowing for objective repeated assessment. The purpose of this study was directly to compare the raw values and reproducibility of results from a novel automated pupillometer smartphone application, MindMirror, against an existing validated pupillometer, the Neuroptics NPi-200. This study was a prospective cohort study comparing pupillary light response values from three consecutive readings from the MindMirror application and the Neuroptics NPi-200 pupillometer in 66 well subjects. Initial diameter values for NPi-200 and MindMirror did not demonstrate a clinically significant mean error (−0.37 mm ± 0.51 mm) with comparable inter-sample repeatability. Other parameters suggestive of reduced constriction amplitude and velocity, were likely to be related to the clinical conditions of collection. The data suggested that the novel MindMirror smartphone application provided accurate results that were predictable and comparable to the gold standard device for automated pupillometry, and may therefore have utility as a portable accessible device for the rapid assessment of pupillary light response, particularly outside the intensive care unit.

Anaesthesia and Intensive Care
Gold Coast Hospital (AU), Clarence Valley Council (AU), Liverpool Hospital (AU), UNSW Sydney (AU), Ingham Institute (AU)
Openalex Percentile: Top 42%
Intraoperative Neuromonitoring and Anesthetic Effects
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A novel pupillometer smartphone application demonstrates accurate results compared with a validated clinical pupillometer: The Pupillary Light Response Validation (PLR-V) Study — William E. Davies, Paul M Middleton, et al. · Anaesthesia and Intensive Care (2025) | TGRS Research Map | TGRS