High precision fluorescence tomography system for pre-clinical radiation research: system design and validation

<i>Objective</i>. Cone-beam computed tomography (CBCT), commonly used for image guidance in pre-clinical studies, is limited in soft tissue localization and lacks the functional information needed for image-guided radiation studies and treatment assessment. To address these limitations, we developed three-dimensional fluorescence tomography (FT) for high precision functional image-guided research and integrated it with small animal irradiators.<i>Methods</i>. The FT system, integrated with a commercial bioluminescence-guided platform in a standalone configuration, enabling compact multi-projection and multi-spectral imaging. A dual-axis galvo mirror scanner was used for laser spots (LSs) scanning to excite internal fluorophore. A transportable mouse bed allows animals imaged in the optical system and transferred to irradiators for CBCT imaging and FT-guided radiation research. Spatial geometry-based methods were developed to map the LSs and fluorescence images onto the animal numerical mesh surface, generated from the CBCT image, to serve as input data for FT reconstruction. A self-calibration method using Born-ratio data, fluorescence normalized to the excitation image, was adopted for reconstruction. Mouse phantom embedded with QD800 reporter and orthotopic glioblastoma (GBM) model injected with IRDye 800CW labeled U87-epidermal growth factor receptor (EGFR) cells were used to verify the FT system in target localization.<i>Results</i>. The LS mapping accuracy are within 0.7 mm maximum deviation. Our system can reconstruct the target QD800 as deep as 17 mm in the mouse phantom with single projection fluorescence imaging at localization accuracy <1 mm deviation. Using the Born ratio approach, we effectively eliminated excitation light leakage and autofluorescence, enabling our FT system to localize IRDye 800CW-labeled EGFR-overexpressing GBM cells in the mouse brain with approximately 1 mm accuracy.<i>Significance</i>. We developed a compact FT system integrated with a small animal irradiator, achieving sub-millimeter localization accuracy in phantom and<i>in vivo</i>models. Its precision, flexibility, and compatibility position it as a powerful tool for advancing functional imaging-guided pre-clinical radiation research.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-12-04
DOI
https://doi.org/10.1088/1361-6560/ae281b
Primary Topic
Optical Imaging and Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High precision fluorescence tomography system for pre-clinical radiation research: system design and validation

Ken Kang‐Hsin Wang, Amyn A. Habib, Hamid Dehghani, Zhishen Tong et al.
Physics in Medicine and Biology
Optical Imaging and Spectroscopy Techniques
article

High precision fluorescence tomography system for pre-clinical radiation research: system design and validation

Ken Kang‐Hsin Wang, Amyn A. Habib, Hamid Dehghani, Zhishen Tong, Iulian Iordachita, Luke A. Hardy, Gao Guo, Xiangkun Xu, John W. Wong, Yu‐Pei Tseng
article en

Abstract

<i>Objective</i>. Cone-beam computed tomography (CBCT), commonly used for image guidance in pre-clinical studies, is limited in soft tissue localization and lacks the functional information needed for image-guided radiation studies and treatment assessment. To address these limitations, we developed three-dimensional fluorescence tomography (FT) for high precision functional image-guided research and integrated it with small animal irradiators.<i>Methods</i>. The FT system, integrated with a commercial bioluminescence-guided platform in a standalone configuration, enabling compact multi-projection and multi-spectral imaging. A dual-axis galvo mirror scanner was used for laser spots (LSs) scanning to excite internal fluorophore. A transportable mouse bed allows animals imaged in the optical system and transferred to irradiators for CBCT imaging and FT-guided radiation research. Spatial geometry-based methods were developed to map the LSs and fluorescence images onto the animal numerical mesh surface, generated from the CBCT image, to serve as input data for FT reconstruction. A self-calibration method using Born-ratio data, fluorescence normalized to the excitation image, was adopted for reconstruction. Mouse phantom embedded with QD800 reporter and orthotopic glioblastoma (GBM) model injected with IRDye 800CW labeled U87-epidermal growth factor receptor (EGFR) cells were used to verify the FT system in target localization.<i>Results</i>. The LS mapping accuracy are within 0.7 mm maximum deviation. Our system can reconstruct the target QD800 as deep as 17 mm in the mouse phantom with single projection fluorescence imaging at localization accuracy <1 mm deviation. Using the Born ratio approach, we effectively eliminated excitation light leakage and autofluorescence, enabling our FT system to localize IRDye 800CW-labeled EGFR-overexpressing GBM cells in the mouse brain with approximately 1 mm accuracy.<i>Significance</i>. We developed a compact FT system integrated with a small animal irradiator, achieving sub-millimeter localization accuracy in phantom and<i>in vivo</i>models. Its precision, flexibility, and compatibility position it as a powerful tool for advancing functional imaging-guided pre-clinical radiation research.

Physics in Medicine and BiologyVol. 70(24)
Johns Hopkins University (US), Johns Hopkins Medicine (US), University of Alabama at Birmingham (US), University of Baltimore (US), University of Birmingham (GB), The University of Texas Southwestern Medical Center (US)
Openalex Percentile: Top 47%
Optical Imaging and Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.