Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study

Augmented whole-body scanning via magnifying PET (AWSM-PET) is a technology that aims to enhance the image resolution and sensitivity of the clinical PET/CT scanner for whole-body imaging. This study presents the system design, performance evaluation, and results from phantoms and initial human imaging studies. <b>Methods:</b> The AWSM-PET system integrates 2 high-resolution detector panels ("outserts") with a Biograph Vision PET/CT scanner (Siemens Healthineers). Positioned outside the scanner's axial field of view, each panel consists of 32 detector modules, each containing a 30 × 30 lutetium oxyorthosilicate crystal array (0.97 × 0.97 × 10 mm³ per crystal) at 1.05-mm pitch. These detectors are packaged and read out using the scanner's detector electronics assembly. Customized firmware and software were developed to establish the additional coincidence detection among the outserts and the scanner. The system acquires data simultaneously during a whole-body scan using continuous bed motion. List-mode-based image reconstruction software has been developed for AWSM-PET system using the continuous-bed-motion protocol. Performance was evaluated through sensitivity measurements, imaging of a mini-resolution phantom containing multiple groups of hot spheres, a National Electrical Manufacturers Association Image Quality (NEMA IQ) phantom study with custom small-diameter tumor inserts, and an initial human study. <b>Results:</b> The outsert detectors achieved an energy resolution of 11% at 511 keV. The coincidence resolving time between the outserts and between outserts and scanner was 183 and 211 ps full width at half maximum, respectively. Sensitivity increased by up to 18.4%, depending on the source location. Mini-resolution phantom images demonstrated higher peak-to-valley ratios for spheres with a diameter of 6 mm or less compared with native scanner images. In the NEMA IQ study, AWSM-PET achieved a higher contrast recovery ratio (CRC) for the smallest spheric lesion (diameter, 4.88 mm) across all CRC metrics (maximum, peak, and mean) but no improvement for larger lesions compared with high-resolution Biograph Vision images. Initial human imaging confirmed the system's compatibility with clinical workflows, achieving improved resolution in high-count regions, such as the brain. However, increased noise in low-count regions (e.g., abdomen) was observed and requires further improvement. <b>Conclusion:</b> The prototype AWSM-PET system demonstrated enhanced image resolution and sensitivity of a clinical PET/CT scanner without negatively impacting its performance, showing improved CRC and spatial resolution for small lesions in phantom and human imaging.

Authors

Institutions

Publication Details

Journal
Journal of Nuclear Medicine
Published
2025-12-04
DOI
https://doi.org/10.2967/jnumed.125.270117
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study

Farrokh Dehdashti, J. Corbeil, Yunlai Chen, Joseph A. O’Sullivan et al.
Journal of Nuclear Medicine
Medical Imaging Techniques and Applications
article

Development of a Prototype AWSM-PET Device for Augmented Whole-Body PET Imaging and Initial Human Study

Farrokh Dehdashti, J. Corbeil, Yunlai Chen, Joseph A. O’Sullivan, Richard Laforest, Darren Pocci, Stefan B. Siegel, Yuan‐Chuan Tai, Robert A. Mintzer, Mahdjoub Hamdi, Samarth Aggarwal, Sergey Komarov, Sanghee Cho, Ling Cai, Anthony J. Thomas, Jennifer B. Frye
article en

Abstract

Augmented whole-body scanning via magnifying PET (AWSM-PET) is a technology that aims to enhance the image resolution and sensitivity of the clinical PET/CT scanner for whole-body imaging. This study presents the system design, performance evaluation, and results from phantoms and initial human imaging studies. <b>Methods:</b> The AWSM-PET system integrates 2 high-resolution detector panels ("outserts") with a Biograph Vision PET/CT scanner (Siemens Healthineers). Positioned outside the scanner's axial field of view, each panel consists of 32 detector modules, each containing a 30 × 30 lutetium oxyorthosilicate crystal array (0.97 × 0.97 × 10 mm³ per crystal) at 1.05-mm pitch. These detectors are packaged and read out using the scanner's detector electronics assembly. Customized firmware and software were developed to establish the additional coincidence detection among the outserts and the scanner. The system acquires data simultaneously during a whole-body scan using continuous bed motion. List-mode-based image reconstruction software has been developed for AWSM-PET system using the continuous-bed-motion protocol. Performance was evaluated through sensitivity measurements, imaging of a mini-resolution phantom containing multiple groups of hot spheres, a National Electrical Manufacturers Association Image Quality (NEMA IQ) phantom study with custom small-diameter tumor inserts, and an initial human study. <b>Results:</b> The outsert detectors achieved an energy resolution of 11% at 511 keV. The coincidence resolving time between the outserts and between outserts and scanner was 183 and 211 ps full width at half maximum, respectively. Sensitivity increased by up to 18.4%, depending on the source location. Mini-resolution phantom images demonstrated higher peak-to-valley ratios for spheres with a diameter of 6 mm or less compared with native scanner images. In the NEMA IQ study, AWSM-PET achieved a higher contrast recovery ratio (CRC) for the smallest spheric lesion (diameter, 4.88 mm) across all CRC metrics (maximum, peak, and mean) but no improvement for larger lesions compared with high-resolution Biograph Vision images. Initial human imaging confirmed the system's compatibility with clinical workflows, achieving improved resolution in high-count regions, such as the brain. However, increased noise in low-count regions (e.g., abdomen) was observed and requires further improvement. <b>Conclusion:</b> The prototype AWSM-PET system demonstrated enhanced image resolution and sensitivity of a clinical PET/CT scanner without negatively impacting its performance, showing improved CRC and spatial resolution for small lesions in phantom and human imaging.

Journal of Nuclear MedicineVol. 67(2)
Washington University in St. Louis (US), Mallinckrodt (United States) (US), Jewish Hospital (US), Siemens Healthcare (United States) (US)
Openalex Percentile: Top 47%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.