Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers

Anaemia is a common finding in critically ill dogs, and understanding its potential impact on the accuracy of point-of-care glucometers is essential. In human medicine, it is well reported that haematocrit can affect blood glucose measurements and influence clinical outcomes when measured using certain hand-held glucometers, whereas evidence in veterinary medicine is still scarce. This prospective case-control study included 72 client-owned dogs, 39 of which were anaemic. Haematocrit values were used to classify dogs into anaemic and non-anaemic categories. Whole blood glucose concentrations were measured using two new models of veterinary hand-held glucometers (G-PET PLUS<sup>®</sup> and AlphaTrak3<sup>®</sup>) and compared with the serum blood glucose measurements obtained with an automated wet biochemistry analyser. Agreement between the point-of-care blood glucose analysers and the reference method, as well as the effects of haematocrit on measurement accuracy, was assessed. Both hand-held glucometers had a significant positive correlation with the wet biochemistry analyser (AlphaTrak3<sup>®</sup>: r = 0.664; 95% CI 0.502, 0.78; G-PET PLUS<sup>®</sup>: r = 0.769; 95% CI 0.643, 0.855). However, when comparing the overall blood glucose results, there was a significant negative percentage bias (<i>p</i> < 0.001) for both devices. Haematocrit reductions did not affect the degree of bias for AlphaTrak3<sup>®</sup>, and for the G-PET PLUS<sup>®</sup>, bias was also present in control patients and in patients with a mild haematocrit reduction. Despite these statistical differences, clinical impact assessments of the results via specialist clinician questionnaires and Parkes error grid analysis showed no change in clinical decision-making or patient outcome, supporting the cautious use of these devices in anaemic but normoglycaemic dogs when testing whole blood EDTA samples.

Authors

Institutions

Publication Details

Journal
Frontiers in Veterinary Science
Published
2025-12-04
DOI
https://doi.org/10.3389/fvets.2025.1694081
Citations
1
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
1.29
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers

Neelakshi Mann, Tim H. Sparks, Helen F. Titmarsh, Montse García et al.
1 citations
Frontiers in Veterinary Science
Hyperglycemia and glycemic control in critically ill and hospitalized patients
1.29
article

Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers

Neelakshi Mann, Tim H. Sparks, Helen F. Titmarsh, Montse García, M. Perelló, Valentina Silic
article en
1 citations

Abstract

Anaemia is a common finding in critically ill dogs, and understanding its potential impact on the accuracy of point-of-care glucometers is essential. In human medicine, it is well reported that haematocrit can affect blood glucose measurements and influence clinical outcomes when measured using certain hand-held glucometers, whereas evidence in veterinary medicine is still scarce. This prospective case-control study included 72 client-owned dogs, 39 of which were anaemic. Haematocrit values were used to classify dogs into anaemic and non-anaemic categories. Whole blood glucose concentrations were measured using two new models of veterinary hand-held glucometers (G-PET PLUS<sup>®</sup> and AlphaTrak3<sup>®</sup>) and compared with the serum blood glucose measurements obtained with an automated wet biochemistry analyser. Agreement between the point-of-care blood glucose analysers and the reference method, as well as the effects of haematocrit on measurement accuracy, was assessed. Both hand-held glucometers had a significant positive correlation with the wet biochemistry analyser (AlphaTrak3<sup>®</sup>: r = 0.664; 95% CI 0.502, 0.78; G-PET PLUS<sup>®</sup>: r = 0.769; 95% CI 0.643, 0.855). However, when comparing the overall blood glucose results, there was a significant negative percentage bias (<i>p</i> < 0.001) for both devices. Haematocrit reductions did not affect the degree of bias for AlphaTrak3<sup>®</sup>, and for the G-PET PLUS<sup>®</sup>, bias was also present in control patients and in patients with a mild haematocrit reduction. Despite these statistical differences, clinical impact assessments of the results via specialist clinician questionnaires and Parkes error grid analysis showed no change in clinical decision-making or patient outcome, supporting the cautious use of these devices in anaemic but normoglycaemic dogs when testing whole blood EDTA samples.

Frontiers in Veterinary ScienceVol. 12
Leicester City Council (GB)
Openalex Percentile: Top 13%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
1.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.