First reported case of multielectrode mapping catheter delamination detected by intracardiac echocardiography during atrial fibrillation ablation

Electroanatomic mapping with multipolar catheters is common practice during atrial fibrillation (AF) ablation. Although these catheters are generally safe, rare complications related to catheter integrity have been described. The use of multipolar mapping catheters may still be associated with specific risks, including thrombus formation, entrapment or injury of valvular structures, and potential mechanical failure because of their complex design with multiple splines and electrodes. We report a case in which intracardiac echocardiography (ICE) enabled early detection of a filamentous foreign body emerging from a PentaRay (Biosense Webster) mapping catheter, likely because of partial delamination of the spline covering.

Authors

Institutions

Publication Details

Journal
HeartRhythm Case Reports
Published
2025-08-15
DOI
https://doi.org/10.1016/j.hrcr.2025.08.009
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

First reported case of multielectrode mapping catheter delamination detected by intracardiac echocardiography during atrial fibrillation ablation

Giuseppe Lumia, Achille Gaspardone, Maria Chiara Gatto, Filippo Lamberti et al.
HeartRhythm Case Reports
Atrial Fibrillation Management and Outcomes
article

First reported case of multielectrode mapping catheter delamination detected by intracardiac echocardiography during atrial fibrillation ablation

Giuseppe Lumia, Achille Gaspardone, Maria Chiara Gatto, Filippo Lamberti, Letizia Verticelli, Cesare Bellini
article en

Abstract

Electroanatomic mapping with multipolar catheters is common practice during atrial fibrillation (AF) ablation. Although these catheters are generally safe, rare complications related to catheter integrity have been described. The use of multipolar mapping catheters may still be associated with specific risks, including thrombus formation, entrapment or injury of valvular structures, and potential mechanical failure because of their complex design with multiple splines and electrodes. We report a case in which intracardiac echocardiography (ICE) enabled early detection of a filamentous foreign body emerging from a PentaRay (Biosense Webster) mapping catheter, likely because of partial delamination of the spline covering.

HeartRhythm Case ReportsVol. 11(11)
St. Eugenio Hospital (IT)
Openalex Percentile: Top 45%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.