Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp

This study quantitatively evaluated the effects of tube voltage, CTDIvol, energy level, and iodine concentration on the performance of lower-extremity vascular imaging using a second-generation dual-layer spectral CT (DSCT). We further assessed the potential for reducing radiation dose and contrast agent use without compromising image quality. A spectral CT phantom with tissue-equivalent inserts and varying iodine concentrations (4, 8, 12, and 16 mgI/mL) was scanned at different tube voltages (100/120 kVp), CTDIvol levels (2.5, 5, 10 mGy), and virtual monoenergetic levels (40, 70, 100 keV), alongside conventional imaging. Vessel diameters of 2.5, 4, and 5 mm were evaluated. Objective image quality was assessed via CT attenuation, SD, SNR, and CNR. Two radiologists used a 5-point Likert scale to assess the subjective image quality. No significant differences were observed in subjective or objective scores between tube voltages or across CTDIvol levels (p > 0.05). Low-energy 40 keV images yielded comparable quality to 70 keV and conventional images. At 40 keV, ultra-low iodine concentrations (4 mgI/mL) achieved image quality similar to higher concentrations (12/16 mgI/mL) in conventional scans. The combination of 100 kVp, 2.5 mGy, and 40 keV provided superior image quality relative to conventional CT. Second-generation DSCT at 100 kVp provides high-accuracy lower-limb vascular imaging, even at low radiation doses and contrast concentrations, supporting its potential for clinical dose-reduction strategies.

Authors

Institutions

Publication Details

Journal
BMC Medical Imaging
Published
2025-12-04
DOI
https://doi.org/10.1186/s12880-025-02047-8
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp

Shenghui Yu, Daming Zhang, Wenjie Zheng, Yun Wang et al.
BMC Medical Imaging
Advanced X-ray and CT Imaging
article

Phantom-based quantification of the spectral accuracy of lower extremity vascular imaging using a second-generation dual-layer spectral CT at 100 kVp and 120kVp

Shenghui Yu, Daming Zhang, Wenjie Zheng, Yun Wang, Yining Wang, Chengqi Zhang, Xiaomei Lü, Xiang Gao, Zehong Lin, Zhengyu Jin, Huadan Xue
article en

Abstract

This study quantitatively evaluated the effects of tube voltage, CTDIvol, energy level, and iodine concentration on the performance of lower-extremity vascular imaging using a second-generation dual-layer spectral CT (DSCT). We further assessed the potential for reducing radiation dose and contrast agent use without compromising image quality. A spectral CT phantom with tissue-equivalent inserts and varying iodine concentrations (4, 8, 12, and 16 mgI/mL) was scanned at different tube voltages (100/120 kVp), CTDIvol levels (2.5, 5, 10 mGy), and virtual monoenergetic levels (40, 70, 100 keV), alongside conventional imaging. Vessel diameters of 2.5, 4, and 5 mm were evaluated. Objective image quality was assessed via CT attenuation, SD, SNR, and CNR. Two radiologists used a 5-point Likert scale to assess the subjective image quality. No significant differences were observed in subjective or objective scores between tube voltages or across CTDIvol levels (p > 0.05). Low-energy 40 keV images yielded comparable quality to 70 keV and conventional images. At 40 keV, ultra-low iodine concentrations (4 mgI/mL) achieved image quality similar to higher concentrations (12/16 mgI/mL) in conventional scans. The combination of 100 kVp, 2.5 mGy, and 40 keV provided superior image quality relative to conventional CT. Second-generation DSCT at 100 kVp provides high-accuracy lower-limb vascular imaging, even at low radiation doses and contrast concentrations, supporting its potential for clinical dose-reduction strategies.

BMC Medical ImagingVol. 25(1)
Harbin University (CN), Philips (Finland) (FI), Harbin Engineering University (CN), Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Peking Union Medical College Hospital (CN)
Peking Union Medical College Hospital, Peking Union Medical College
Openalex Percentile: Top 62%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.