Performance of low-coherence and confocal refractometry with reduced index contrast

This paper looks to evaluate the performance of combined low-coherence and confocal refractometry under conditions of reduced refractive index contrast. The instrument measures the refractive index and thickness of transparent objects using a fibre-based low-coherence interferometer with a line-scan spectrometer. A sample was designed that mimics the on-axis structure of the eye, consisting of a lens tube holding a pair of windows surrounded on either side by fluid chambers and sealed at the far end by a diffuse black metallic plug. Sucrose solutions with a range of concentrations were injected into the fluid chambers, providing a linear variation in refractive index of 1.3330 to 1.4416. The instrument was used to simultaneously measure both the phase and group refractive indices <i>n</i><sub>p</sub> and <i>n</i><sub>g</sub>, as well as the physical thickness <i>t</i> of the windows and the fluid in the chambers. Although the measurement accuracy is shown to decrease with reduced refractive index contrast, it remains better than 0.6% over all measured components for sucrose concentrations of up to 60.3%, which is close to the saturation limit.

Authors

Publication Details

Journal
Optics Express
Published
2025-12-04
DOI
https://doi.org/10.1364/oe.582088
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of low-coherence and confocal refractometry with reduced index contrast

J. M. Hallam, Ralph P. Tatam, Daniel Francis
Optics Express
Ophthalmology and Visual Impairment Studies
article

Performance of low-coherence and confocal refractometry with reduced index contrast

J. M. Hallam, Ralph P. Tatam, Daniel Francis
article en

Abstract

This paper looks to evaluate the performance of combined low-coherence and confocal refractometry under conditions of reduced refractive index contrast. The instrument measures the refractive index and thickness of transparent objects using a fibre-based low-coherence interferometer with a line-scan spectrometer. A sample was designed that mimics the on-axis structure of the eye, consisting of a lens tube holding a pair of windows surrounded on either side by fluid chambers and sealed at the far end by a diffuse black metallic plug. Sucrose solutions with a range of concentrations were injected into the fluid chambers, providing a linear variation in refractive index of 1.3330 to 1.4416. The instrument was used to simultaneously measure both the phase and group refractive indices <i>n</i><sub>p</sub> and <i>n</i><sub>g</sub>, as well as the physical thickness <i>t</i> of the windows and the fluid in the chambers. Although the measurement accuracy is shown to decrease with reduced refractive index contrast, it remains better than 0.6% over all measured components for sucrose concentrations of up to 60.3%, which is close to the saturation limit.

Optics ExpressVol. 34(2)
Openalex Percentile: Top 50%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.