Acute Thrombosis During Atrial Leadless Pacemaker Implantation
Aveir AR atrial LP implantation can be complicated by acute thrombosis, particularly in hypercoagulable state and prolonged procedural time. Acute thrombus formation may manifest as abnormally high impedance during LP mapping.
Authors
- Michael R. Gold (ORCID: https://orcid.org/0000-0002-4579-0216)
- Yu Jiang (ORCID: https://orcid.org/0000-0002-2797-1099)
- Yiran Hu (ORCID: https://orcid.org/0000-0003-3429-4003)
- Hongxia Niu (ORCID: https://orcid.org/0000-0003-3033-4741)
- Wei Hua (ORCID: https://orcid.org/0000-0001-6921-9375)
- Jie Yang (ORCID: https://orcid.org/0000-0001-7661-2861)
- Bing Fu (ORCID: https://orcid.org/0000-0003-4734-2457)
- Xiaohui Ning (ORCID: https://orcid.org/0000-0002-5373-7511)
Institutions
- Medical University of South Carolina (US)
- Chinese Academy of Medical Sciences & Peking Union Medical College (CN)
- Fu Wai Hospital (CN)
Publication Details
- Journal
- JACC Case Reports
- Published
- 2025-12-01
- DOI
- https://doi.org/10.1016/j.jaccas.2025.106152
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00