In Vivo Glx Measurements From GABA‐Edited HERMES at 3 T Are Not Consistent With Those From Short‐TE PRESS Across Scanners, Brain Regions, Diagnostic and Age Groups

1H-Magnetic resonance spectroscopy (1H-MRS) is a noninvasive technique for quantifying brain metabolites, including glutamate, glutathione (GSH), and γ-aminobutyric acid (GABA), which are essential for brain function and implicated in various neurodevelopmental conditions. As such, 1H-MRS methods that enable reliable and accurate measurement of these metabolites are of considerable clinical value. Hadamard Encoding and Reconstruction of MEGA-Edited Spectroscopy (HERMES; echo time [TE] = 80 ms) is a spectral editing technique that allows for the simultaneous quantification of GABA and GSH, using subtraction approaches to resolve these metabolites in a difference spectrum. Additionally, glutamate plus glutamine resonances (Glx) can be resolved either from the HERMES GABA-edited difference spectrum (GABA-DIFF) or from the sum of all HERMES transients (SUM spectrum). However, the reliability of 80-ms HERMES for quantification of Glx has not been systematically assessed. Here, we evaluate the agreement between Glx obtained from HERMES GABA-DIFF and SUM spectra with Glx derived from short-TE PRESS (TE = 35 ms), which is conventionally used for Glx estimation and has demonstrated reproducibility. Data were acquired from 139 participants across two brain regions (ACC and Thalamus voxels), three scanners, two diagnostic groups (autism and neurotypical development) and two age groups (adolescent/adult and preschooler). Comparisons were made using both creatine-scaled and tissue-corrected Glx estimates. Our findings demonstrate significant systematic and proportional bias between Glx estimates from HERMES (SUM and GABA-DIFF) and short-TE PRESS, consistent across scanners, voxels, age groups and diagnostic categories. These findings indicate that Glx estimates derived from HERMES are not directly comparable to those from short-TE PRESS, and this discrepancy is consistent across a multisite study setting. This underscores the importance of sequence selection and careful methodological consideration when integrating and interpreting data from 1H-MRS across different acquisition protocols.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-12-03
DOI
https://doi.org/10.1002/nbm.70171
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Vivo Glx Measurements From GABA‐Edited HERMES at 3 T Are Not Consistent With Those From Short‐TE PRESS Across Scanners, Brain Regions, Diagnostic and Age Groups

Alice R. Thomson, Isabel Yorke, Herbert Roeyers, Laura Colomar et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

In Vivo Glx Measurements From GABA‐Edited HERMES at 3 T Are Not Consistent With Those From Short‐TE PRESS Across Scanners, Brain Regions, Diagnostic and Age Groups

Alice R. Thomson, Isabel Yorke, Herbert Roeyers, Laura Colomar, Giorgia Bussu, Tony Charman, Amy Goodwin, Declan Murphy, Ellen Demurie, Pascal-M. Aggensteiner, Pim Pullens, Bethany Oakley, Tomoki Arichi, Viola Hollestein, Manon A. Krol, Nicolaas A.J. Puts, Anne Fritz, Rosemary Holt, Terje Falck‐Ytter, Eva Loth, Edward Bullock, Nele Herregods, Jan K. Buitelaar, Matthew Bowdler, Julia Koziel
article en

Abstract

1H-Magnetic resonance spectroscopy (1H-MRS) is a noninvasive technique for quantifying brain metabolites, including glutamate, glutathione (GSH), and γ-aminobutyric acid (GABA), which are essential for brain function and implicated in various neurodevelopmental conditions. As such, 1H-MRS methods that enable reliable and accurate measurement of these metabolites are of considerable clinical value. Hadamard Encoding and Reconstruction of MEGA-Edited Spectroscopy (HERMES; echo time [TE] = 80 ms) is a spectral editing technique that allows for the simultaneous quantification of GABA and GSH, using subtraction approaches to resolve these metabolites in a difference spectrum. Additionally, glutamate plus glutamine resonances (Glx) can be resolved either from the HERMES GABA-edited difference spectrum (GABA-DIFF) or from the sum of all HERMES transients (SUM spectrum). However, the reliability of 80-ms HERMES for quantification of Glx has not been systematically assessed. Here, we evaluate the agreement between Glx obtained from HERMES GABA-DIFF and SUM spectra with Glx derived from short-TE PRESS (TE = 35 ms), which is conventionally used for Glx estimation and has demonstrated reproducibility. Data were acquired from 139 participants across two brain regions (ACC and Thalamus voxels), three scanners, two diagnostic groups (autism and neurotypical development) and two age groups (adolescent/adult and preschooler). Comparisons were made using both creatine-scaled and tissue-corrected Glx estimates. Our findings demonstrate significant systematic and proportional bias between Glx estimates from HERMES (SUM and GABA-DIFF) and short-TE PRESS, consistent across scanners, voxels, age groups and diagnostic categories. These findings indicate that Glx estimates derived from HERMES are not directly comparable to those from short-TE PRESS, and this discrepancy is consistent across a multisite study setting. This underscores the importance of sequence selection and careful methodological consideration when integrating and interpreting data from 1H-MRS across different acquisition protocols.

NMR in BiomedicineVol. 39(1)
Uppsala University (SE), HOGENT University of Applied Sciences and Arts (BE), Radboud University Nijmegen (NL), King's College London (GB), University of Cambridge (GB), Ghent University Hospital (BE), Radboud University Medical Center (NL), Karolinska Institutet (SE), Central Institute of Mental Health (DE), Stockholm Health Care Services (SE)
Autism Speaks, South London and Maudsley NHS Foundation Trust, European Federation of Pharmaceutical Industries and Associations, Simons Foundation Autism Research Initiative, National Institute for Health and Care Research, King's College London, European Commission, Medical Research Council Centre for Neurodevelopmental Disorders, NIHR Maudsley Biomedical Research Centre, Medical Research Council
Openalex Percentile: Top 47%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.