The In Vitro Biomechanics of Locking Plates Used in a Supercutaneous Fashion Compared to a Type Ia External Skeletal Fixator

The aim of this study was to determine the biomechanics of a supercutaneously placed String of Pearls plate (SOP) and supercutaneously placed locking compression plate (LCP) to be used in lieu of a type Ia external fixator.Nine surrogate constructs of 3.5 mm SOP, 3.5 mm LCP and type Ia external fixator offset 20 mm from the cortex over a 10-mm fracture gap were tested in non-destructive loading to 200 N and then tested in load to 100% displacement.Maximum displacement was used to calculate median maximum displacement in non-destructive axial compression testing for the SOP, LCP and external fixator, measuring 8.5 mm (interquartile range [IQR] 8.23-8.95), 10.3 mm (IQR 9.99-10.40) and 1.2 mm (IQR 1.17-1.31), respectively. There was a difference between the external fixator to LCP (<i>p</i> = 0.0050), external fixator to SOP (<i>p</i> = 0.0050) and LCP to SOP (<i>p</i> = 0.0050). Maximum load was used in compression to 100% displacement to calculate the median maximum load. The SOP, LCP and external fixator, withstood median loads to 100% displacement of 244 N (IQR 231.83-257.19), 168 N (IQR 165.07-170.86) and 766 N (IQR 688.14-784.29), respectively. There was a difference between external fixator to SOP (<i>p</i> = 0.0047), external fixator to LCP (<i>p</i> ≤ 0.001) and LCP to SOP (<i>p</i> = 0.0024).The type Ia external fixator is significantly stiffer compared with the supercutaneous 3.5 mm SOP and 3.5 mm LCP. Given the low stiffness and high displacement, supercutaneously placed locking plates may be unsuitable for clinical fracture repair in a 20-kg dog with an expected axial load of 200 N.

Authors

Institutions

Publication Details

Journal
Veterinary and Comparative Orthopaedics and Traumatology
Published
2025-12-02
DOI
https://doi.org/10.1055/a-2738-5771
Primary Topic
Bone fractures and treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The In Vitro Biomechanics of Locking Plates Used in a Supercutaneous Fashion Compared to a Type Ia External Skeletal Fixator

Gabrielle Monteith, John Runciman, Ashley R. Proctor, Melissa A. MacIver et al.
Veterinary and Comparative Orthopaedics and Traumatology
Bone fractures and treatments
article

The In Vitro Biomechanics of Locking Plates Used in a Supercutaneous Fashion Compared to a Type Ia External Skeletal Fixator

Gabrielle Monteith, John Runciman, Ashley R. Proctor, Melissa A. MacIver, Scott C.E. Brandon, Noël Moens
article en

Abstract

The aim of this study was to determine the biomechanics of a supercutaneously placed String of Pearls plate (SOP) and supercutaneously placed locking compression plate (LCP) to be used in lieu of a type Ia external fixator.Nine surrogate constructs of 3.5 mm SOP, 3.5 mm LCP and type Ia external fixator offset 20 mm from the cortex over a 10-mm fracture gap were tested in non-destructive loading to 200 N and then tested in load to 100% displacement.Maximum displacement was used to calculate median maximum displacement in non-destructive axial compression testing for the SOP, LCP and external fixator, measuring 8.5 mm (interquartile range [IQR] 8.23-8.95), 10.3 mm (IQR 9.99-10.40) and 1.2 mm (IQR 1.17-1.31), respectively. There was a difference between the external fixator to LCP (<i>p</i> = 0.0050), external fixator to SOP (<i>p</i> = 0.0050) and LCP to SOP (<i>p</i> = 0.0050). Maximum load was used in compression to 100% displacement to calculate the median maximum load. The SOP, LCP and external fixator, withstood median loads to 100% displacement of 244 N (IQR 231.83-257.19), 168 N (IQR 165.07-170.86) and 766 N (IQR 688.14-784.29), respectively. There was a difference between external fixator to SOP (<i>p</i> = 0.0047), external fixator to LCP (<i>p</i> ≤ 0.001) and LCP to SOP (<i>p</i> = 0.0024).The type Ia external fixator is significantly stiffer compared with the supercutaneous 3.5 mm SOP and 3.5 mm LCP. Given the low stiffness and high displacement, supercutaneously placed locking plates may be unsuitable for clinical fracture repair in a 20-kg dog with an expected axial load of 200 N.

Veterinary and Comparative Orthopaedics and Traumatology
University of Guelph (CA)
Openalex Percentile: Top 51%
Bone fractures and treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

The In Vitro Biomechanics of Locking Plates Used in a Supercutaneous Fashion Compared to a Type Ia External Skeletal Fixator — Gabrielle Monteith, John Runciman, et al. · Veterinary and Comparative Orthopaedics and Traumatology (2025) | TGRS Research Map | TGRS