Fast operating room scattered radiation calculation in x-ray guided interventions by using deep learning

Protecting medical personnel from the harmful effects of scattered ionising radiation during x-ray-guided procedures is a critical concern. Due to the complex and invisible nature of x-rays, monitoring radiation exposure has been challenging. Existing real-time dosimeters have shown low accuracy and practical limitations. To address these challenges, this study introduces an innovative approach that combines Monte Carlo (MC) simulations and deep learning (DL) for real-time estimation of three-dimensional (3D) scattered radiation in the operating room. The neural network was trained to map patient morphology and imaging parameters to radiation maps, allowing it to adapt to various clinical scenarios. The results demonstrate that the system showcases exceptional speed by efficiently computing 3D radiation maps in 11 ms using modern GPU (NVIDIA RTX 2080). Validation experiments confirmed the reliability of the predicted scatter maps, with a mean absolute percentage error of 10.97% relative to MC simulations. When used to compute organ doses via voxelised-source simulations, the global average organ dose error was 8.2 ± 4.1%. Therefore, the combination of MC simulations and DL provides a promising solution for enhancing the safety of medical personnel during x-ray-guided procedures.

Authors

Institutions

Publication Details

Journal
Journal of Radiological Protection
Published
2025-12-01
DOI
https://doi.org/10.1088/1361-6498/ae268e
Citations
1
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
1.88
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fast operating room scattered radiation calculation in x-ray guided interventions by using deep learning

Mateo Villa, Chi-Hieu Pham, Hussein Harb, Didier Benoit et al.
1 citations
Journal of Radiological Protection
Advanced Radiotherapy Techniques
1.88
article

Fast operating room scattered radiation calculation in x-ray guided interventions by using deep learning

Mateo Villa, Chi-Hieu Pham, Hussein Harb, Didier Benoit, Bahaa Nasr, Julien Bert
article en
1 citations

Abstract

Protecting medical personnel from the harmful effects of scattered ionising radiation during x-ray-guided procedures is a critical concern. Due to the complex and invisible nature of x-rays, monitoring radiation exposure has been challenging. Existing real-time dosimeters have shown low accuracy and practical limitations. To address these challenges, this study introduces an innovative approach that combines Monte Carlo (MC) simulations and deep learning (DL) for real-time estimation of three-dimensional (3D) scattered radiation in the operating room. The neural network was trained to map patient morphology and imaging parameters to radiation maps, allowing it to adapt to various clinical scenarios. The results demonstrate that the system showcases exceptional speed by efficiently computing 3D radiation maps in 11 ms using modern GPU (NVIDIA RTX 2080). Validation experiments confirmed the reliability of the predicted scatter maps, with a mean absolute percentage error of 10.97% relative to MC simulations. When used to compute organ doses via voxelised-source simulations, the global average organ dose error was 8.2 ± 4.1%. Therefore, the combination of MC simulations and DL provides a promising solution for enhancing the safety of medical personnel during x-ray-guided procedures.

Journal of Radiological ProtectionVol. 45(4)
Inserm (FR), Université de Bretagne Occidentale (FR), Centre Hospitalier Régional Universitaire de Brest (FR), Laboratoire de Traitement de l'Information Médicale (FR)
Openalex Percentile: Top 11%
Advanced Radiotherapy Techniques
1.88
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Fast operating room scattered radiation calculation in x-ray guided interventions by using deep learning — Mateo Villa, Chi-Hieu Pham, et al. · Journal of Radiological Protection (2025) | TGRS Research Map | TGRS