When the All-Purpose Tool Fails: Manufacturer-Specific Time Dependency of Magnet Mode in Cardiac Implantable Electronic Devices

<b>Background:</b> Previously unreported, the induction of the magnet mode is time-dependent, according to the cardiac implantable electronic device (CIED) manufacturer, directly after device interrogation. The aim of this study was to systematically investigate the response of CIED from all major manufacturers to the application of a magnet. <b>Methods:</b> CIED from all manufacturers were utilized and connected to an interactive heart simulator (InterSim III, IB Lang). After the end of CIED interrogation, a CIED magnet was placed over the device, and the response was analyzed. <b>Results:</b> Fifteen ICD and eight pacemakers were included. ICDs from the manufacturers Abbott, Boston Scientific, Medtronic and Microport reacted immediately to magnet application by inhibiting antitachycardia function directly after interrogation. In the Biotronik ICD, the magnet mode was only inducible five to seven minutes after the end of the interrogation. In addition, after eight hours of magnet application, the antitachycardia function was automatically and permanently reactivated in all Biotronik ICDs. Pacemakers of Biotronik, Abbott, Boston Scientific, and Microport responded immediately after device interrogation regarding the magnet application. In contrast, Medtronic pacemakers responded only 1.5 min after device interrogation. <b>Conclusions:</b> Magnet mode induction directly after CIED interrogation is manufacturer-specific. Our findings might be of importance when performing invasive procedures with devices that cause electrical interference and in palliative care.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-11-29
DOI
https://doi.org/10.3390/diagnostics15233056
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

When the All-Purpose Tool Fails: Manufacturer-Specific Time Dependency of Magnet Mode in Cardiac Implantable Electronic Devices

Julian Wolfes, Lars Eckardt, Gerrit Frommeyer, Fabienne Kreimer et al.
Diagnostics
Cardiac pacing and defibrillation studies
article

When the All-Purpose Tool Fails: Manufacturer-Specific Time Dependency of Magnet Mode in Cardiac Implantable Electronic Devices

Julian Wolfes, Lars Eckardt, Gerrit Frommeyer, Fabienne Kreimer, Dennis Korthals, Felix K. Wegner, Florian Reinke, Florian Doldi, Christian Ellermann
article en

Abstract

<b>Background:</b> Previously unreported, the induction of the magnet mode is time-dependent, according to the cardiac implantable electronic device (CIED) manufacturer, directly after device interrogation. The aim of this study was to systematically investigate the response of CIED from all major manufacturers to the application of a magnet. <b>Methods:</b> CIED from all manufacturers were utilized and connected to an interactive heart simulator (InterSim III, IB Lang). After the end of CIED interrogation, a CIED magnet was placed over the device, and the response was analyzed. <b>Results:</b> Fifteen ICD and eight pacemakers were included. ICDs from the manufacturers Abbott, Boston Scientific, Medtronic and Microport reacted immediately to magnet application by inhibiting antitachycardia function directly after interrogation. In the Biotronik ICD, the magnet mode was only inducible five to seven minutes after the end of the interrogation. In addition, after eight hours of magnet application, the antitachycardia function was automatically and permanently reactivated in all Biotronik ICDs. Pacemakers of Biotronik, Abbott, Boston Scientific, and Microport responded immediately after device interrogation regarding the magnet application. In contrast, Medtronic pacemakers responded only 1.5 min after device interrogation. <b>Conclusions:</b> Magnet mode induction directly after CIED interrogation is manufacturer-specific. Our findings might be of importance when performing invasive procedures with devices that cause electrical interference and in palliative care.

DiagnosticsVol. 15(23)
University Hospital Münster (DE)
Openalex Percentile: Top 45%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.