The Effect of Lesion Size and Activity Concentration on Image Quality in PET/CT Imaging: A Phantom Study

In this study, the National Electrical Manufacturers Association (NEMA) image quality tests were conducted using the General Electric Discovery 690 model — a hybrid system of positron emission tomography/computed tomography. Quantitative evaluation of image quality within this model primarily involves the use of parameters such as contrast (C), contrast-to-noise ratio (CNR), and contrast recovery coefficient (CRC). These metrics are considered essential performance indicators, particularly in standardized assessment protocols such as those proposed by NEMA. The objective of this study was to evaluate image quality in a NEMA positron emission tomography phantom using 18F-FDG at different signal-to-background (S/Bg) ratios and varying activity concentrations, and to determine the minimum detectable lesion size. Positron emission tomography images were acquired using a NEMA 2012/IEC 2008 phantom with 18F-FDG at S/Bg ratios of 4:1, 6:1, and 8:1, across decreasing activity concentrations ranging from 9.98 to 1.93 MBq/kg. At an S/Bg ratio of 4:1, activity concentrations below 3.90 MBq/kg (as well as at 6:1 below 2.76 MBq/kg) resulted in CNR values less than 5, falling below the detectability threshold. Regarding the CRC [%] values, for the 4:1 ratio, lesions with diameters of 13 and 10 mm fell outside the acceptable limits at concentrations below 2.89 MBq/kg. Similarly, for the 6:1 ratio, lesions with a diameter of 10 mm exceeded these limits at concentrations below 5.29 MBq/kg. Multiple acquisitions were performed at different ratios and activity levels, and the lesion detectability for each condition was determined based on the Rose criterion (CNR ≥ 5).

Authors

Institutions

Publication Details

Journal
Acta Physica Polonica A
Published
2025-12-02
DOI
https://doi.org/10.12693/aphyspola.148.220
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Effect of Lesion Size and Activity Concentration on Image Quality in PET/CT Imaging: A Phantom Study

Serkan İşgören, Gözde Dağlıöz Görür, C. Yalçın, Nami Yeyin et al.
Acta Physica Polonica A
Medical Imaging Techniques and Applications
article

The Effect of Lesion Size and Activity Concentration on Image Quality in PET/CT Imaging: A Phantom Study

Serkan İşgören, Gözde Dağlıöz Görür, C. Yalçın, Nami Yeyin, H.O. Tekın, R. T. Güray, Mustafa Demır, A. Karayel, Hakan Demir
article en

Abstract

In this study, the National Electrical Manufacturers Association (NEMA) image quality tests were conducted using the General Electric Discovery 690 model — a hybrid system of positron emission tomography/computed tomography. Quantitative evaluation of image quality within this model primarily involves the use of parameters such as contrast (C), contrast-to-noise ratio (CNR), and contrast recovery coefficient (CRC). These metrics are considered essential performance indicators, particularly in standardized assessment protocols such as those proposed by NEMA. The objective of this study was to evaluate image quality in a NEMA positron emission tomography phantom using 18F-FDG at different signal-to-background (S/Bg) ratios and varying activity concentrations, and to determine the minimum detectable lesion size. Positron emission tomography images were acquired using a NEMA 2012/IEC 2008 phantom with 18F-FDG at S/Bg ratios of 4:1, 6:1, and 8:1, across decreasing activity concentrations ranging from 9.98 to 1.93 MBq/kg. At an S/Bg ratio of 4:1, activity concentrations below 3.90 MBq/kg (as well as at 6:1 below 2.76 MBq/kg) resulted in CNR values less than 5, falling below the detectability threshold. Regarding the CRC [%] values, for the 4:1 ratio, lesions with diameters of 13 and 10 mm fell outside the acceptable limits at concentrations below 2.89 MBq/kg. Similarly, for the 6:1 ratio, lesions with a diameter of 10 mm exceeded these limits at concentrations below 5.29 MBq/kg. Multiple acquisitions were performed at different ratios and activity levels, and the lesion detectability for each condition was determined based on the Rose criterion (CNR ≥ 5).

Acta Physica Polonica AVol. 148(3)
Istanbul University-Cerrahpaşa (TR), Istinye University (TR), Kocaeli Üniversitesi (TR), Istanbul University (TR)
Openalex Percentile: Top 47%
Medical Imaging Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.