Use of Prussian Blue pseudocapacitive properties to amplify the pulsed amperometric readout of biosensors

Enzymatic electrochemical biosensors are a cornerstone technology in enabling further advancements in the field of Continuous Glucose Monitoring (CGM). Pulsed amperometric methods improve the sensitivity and accuracy of electrochemical biosensors. The literature shows that pulsed amperometry increases the sensitivity of enzymatic glucose biosensors based on Prussian Blue (PB). However, the underlying mechanism responsible for this improvement is poorly understood, which impedes further development of this promising measurement method. The present work elucidates the role of the spontaneous reaction between hydrogen peroxide (H<sub>2</sub>O<sub>2</sub>) and Prussian White (PW) in the sensitivity improvement observed with pulsed amperometry. A charged working electrode (WE) containing PW can catalyze the H<sub>2</sub>O<sub>2</sub> reduction in the open-circuit regime (OCP). The consumption of H<sub>2</sub>O<sub>2</sub> over a 30-min contact at OCP was 65 % at a PW WE, compared to 13 % at a PB WE. This spontaneous process is associated with a partial discharge of the WE (PW → PB) between the amperometric pulses. The subsequent re-charging (PB → PW) yields the current amplification observed with pulsed amperometry. Based on this consideration, we developed and validated a model for glucose quantification using pulsed amperometry that considers the spontaneous reaction of H<sub>2</sub>O<sub>2</sub> with PW. The model achieves 0.998 determination coefficient between glucose concentration and four analytical signals. The insights presented in this work support the optimization and development of the pulsed amperometric detection method in enzymatic glucose biosensors. Additionally, this work advances the understanding of H<sub>2</sub>O<sub>2</sub> detection at PB-based sensors, and contributes to the development of precise and accurate enzymatic glucose biosensors.

Authors

Institutions

Publication Details

Journal
Biosensors and Bioelectronics
Published
2025-12-01
DOI
https://doi.org/10.1016/j.bios.2025.118283
Primary Topic
Electrochemical sensors and biosensors
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Use of Prussian Blue pseudocapacitive properties to amplify the pulsed amperometric readout of biosensors

Johan Bobacka, Tommi Palomäki, Alejandro García Pérez, Eszter Supala et al.
Biosensors and Bioelectronics
Electrochemical sensors and biosensors
article

Use of Prussian Blue pseudocapacitive properties to amplify the pulsed amperometric readout of biosensors

Johan Bobacka, Tommi Palomäki, Alejandro García Pérez, Eszter Supala, Edward Hæggström, Luca Guagneli
article en

Abstract

Enzymatic electrochemical biosensors are a cornerstone technology in enabling further advancements in the field of Continuous Glucose Monitoring (CGM). Pulsed amperometric methods improve the sensitivity and accuracy of electrochemical biosensors. The literature shows that pulsed amperometry increases the sensitivity of enzymatic glucose biosensors based on Prussian Blue (PB). However, the underlying mechanism responsible for this improvement is poorly understood, which impedes further development of this promising measurement method. The present work elucidates the role of the spontaneous reaction between hydrogen peroxide (H<sub>2</sub>O<sub>2</sub>) and Prussian White (PW) in the sensitivity improvement observed with pulsed amperometry. A charged working electrode (WE) containing PW can catalyze the H<sub>2</sub>O<sub>2</sub> reduction in the open-circuit regime (OCP). The consumption of H<sub>2</sub>O<sub>2</sub> over a 30-min contact at OCP was 65 % at a PW WE, compared to 13 % at a PB WE. This spontaneous process is associated with a partial discharge of the WE (PW → PB) between the amperometric pulses. The subsequent re-charging (PB → PW) yields the current amplification observed with pulsed amperometry. Based on this consideration, we developed and validated a model for glucose quantification using pulsed amperometry that considers the spontaneous reaction of H<sub>2</sub>O<sub>2</sub> with PW. The model achieves 0.998 determination coefficient between glucose concentration and four analytical signals. The insights presented in this work support the optimization and development of the pulsed amperometric detection method in enzymatic glucose biosensors. Additionally, this work advances the understanding of H<sub>2</sub>O<sub>2</sub> detection at PB-based sensors, and contributes to the development of precise and accurate enzymatic glucose biosensors.

Biosensors and BioelectronicsVol. 295
Åbo Akademi University (FI), University of Helsinki (FI)
Openalex Percentile: Top 62%
Electrochemical sensors and biosensors
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.