Feasibility and efficacy of a thermal ablation device utilizing a nitinol-based coil

BackgroundA fibrous actuator, composed of a nitinol-based coil memory alloy, functions as a flexible coil under normal conditions. It contracts and generates heat when activated with an electric current. This property makes it a promising candidate for use as an interventional thermal device.PurposeTo evaluate the thermal effects of a nitinol coil ablation device on liver tissue and the intrahepatic portal vein.Material and MethodsAn in vitro experiment using bovine resected liver was performed. Ten Japanese white rabbits were divided equally into two groups (0.6-W and 1.1-W) based on electric power settings. Ablation was performed in three liver regions for each rabbit. Ablated areas were measured using fat-suppressed T1-weighted imaging following gadolinium administration and histological examination with hematoxylin and eosin staining. For the intrahepatic portal vein ablation, nine rabbits were assigned to the ablation group and eight rabbits served as controls. Portography was conducted before and after the procedure.ResultsIn vitro, temperature increases at ablation site and at 5-, 10-, and 15-mm distances and ellipsoid-shaped ablated area were proportional to electric power. In the in vivo experiment, the 1.1-W group exhibited significantly greater ablation effects compared to the 0.6-W group, as measured by magnetic resonance imaging (50.6 vs. 22.7 mm²; <i>P</i> <0.001) and histopathology (38.9 vs. 21.7 mm²; <i>P</i> = 0.001). Nearly complete occlusion (>80% stenosis) was more frequently observed in the ablation group (72.2% vs. 22.2%; <i>P</i> = 0.007).ConclusionThe thermal ablation device utilizing a nitinol coil demonstrated effective liver ablation. Furthermore, it shows potential for customization in endovascular ablation applications.

Authors

Institutions

Publication Details

Journal
Acta Radiologica
Published
2025-12-01
DOI
https://doi.org/10.1177/02841851251396571
Primary Topic
Hepatocellular Carcinoma Treatment and Prognosis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility and efficacy of a thermal ablation device utilizing a nitinol-based coil

Yugo Imai, Yoshiyuki Watanabe, Akinaga Sonoda, Shohei Chatani et al.
Acta Radiologica
Hepatocellular Carcinoma Treatment and Prognosis
article

Feasibility and efficacy of a thermal ablation device utilizing a nitinol-based coil

Yugo Imai, Yoshiyuki Watanabe, Akinaga Sonoda, Shohei Chatani, Ken‐ichi Mukaisho, Akitoshi Inoue, Norihisa Nitta, Naotake Nakamura, Fuyuki Nagano
article en

Abstract

BackgroundA fibrous actuator, composed of a nitinol-based coil memory alloy, functions as a flexible coil under normal conditions. It contracts and generates heat when activated with an electric current. This property makes it a promising candidate for use as an interventional thermal device.PurposeTo evaluate the thermal effects of a nitinol coil ablation device on liver tissue and the intrahepatic portal vein.Material and MethodsAn in vitro experiment using bovine resected liver was performed. Ten Japanese white rabbits were divided equally into two groups (0.6-W and 1.1-W) based on electric power settings. Ablation was performed in three liver regions for each rabbit. Ablated areas were measured using fat-suppressed T1-weighted imaging following gadolinium administration and histological examination with hematoxylin and eosin staining. For the intrahepatic portal vein ablation, nine rabbits were assigned to the ablation group and eight rabbits served as controls. Portography was conducted before and after the procedure.ResultsIn vitro, temperature increases at ablation site and at 5-, 10-, and 15-mm distances and ellipsoid-shaped ablated area were proportional to electric power. In the in vivo experiment, the 1.1-W group exhibited significantly greater ablation effects compared to the 0.6-W group, as measured by magnetic resonance imaging (50.6 vs. 22.7 mm²; <i>P</i> <0.001) and histopathology (38.9 vs. 21.7 mm²; <i>P</i> = 0.001). Nearly complete occlusion (>80% stenosis) was more frequently observed in the ablation group (72.2% vs. 22.2%; <i>P</i> = 0.007).ConclusionThe thermal ablation device utilizing a nitinol coil demonstrated effective liver ablation. Furthermore, it shows potential for customization in endovascular ablation applications.

Acta RadiologicaVol. 67(2)
Ritsumeikan University (JP), Shiga University of Medical Science (JP), Shiga University (JP)
Openalex Percentile: Top 51%
Hepatocellular Carcinoma Treatment and Prognosis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.