Design, Modeling, and Testing—A Compact Variable-Stiffness Actuator for Knee Joint Dimensions

Stroke rehabilitation exoskeletons require joint mechanisms capable of replicating physiological stiffness modulation to adapt to varying gait phases. This paper presents a novel compact variable-stiffness mechanism (VSM) for knee exoskeletons, based on a simplified three-bar linkage topology. The proposed design achieves a pre-configurable quasi-stiffness range of 0.15-2.0 NM/deg. Static characterization under a 2 kg load demonstrated up to 23.0 N of collision force attenuation in softening regimes (λ < 2.3) through passive viscoelastic dissipation, whereas hardening behavior (λ ≥ 2.3) preserved precise torque-angle characteristics scalable to physiological loading. A parametric analysis showed an 89% correlation between the theoretical and scaled experimental stiffness profiles for values from 0.5 to 2.5. The proposed architecture enables decoupled optimization of impact safety and positional precision, offering a clinically adaptable solution for hemiparetic gait assistance.

Authors

Institutions

Publication Details

Journal
Micromachines
Published
2025-11-29
DOI
https://doi.org/10.3390/mi16121365
Primary Topic
Prosthetics and Rehabilitation Robotics
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design, Modeling, and Testing—A Compact Variable-Stiffness Actuator for Knee Joint Dimensions

Guoning Si, Zhuo Zhang, Zilong Guo
Micromachines
Prosthetics and Rehabilitation Robotics
article

Design, Modeling, and Testing—A Compact Variable-Stiffness Actuator for Knee Joint Dimensions

Guoning Si, Zhuo Zhang, Zilong Guo
article en

Abstract

Stroke rehabilitation exoskeletons require joint mechanisms capable of replicating physiological stiffness modulation to adapt to varying gait phases. This paper presents a novel compact variable-stiffness mechanism (VSM) for knee exoskeletons, based on a simplified three-bar linkage topology. The proposed design achieves a pre-configurable quasi-stiffness range of 0.15-2.0 NM/deg. Static characterization under a 2 kg load demonstrated up to 23.0 N of collision force attenuation in softening regimes (λ < 2.3) through passive viscoelastic dissipation, whereas hardening behavior (λ ≥ 2.3) preserved precise torque-angle characteristics scalable to physiological loading. A parametric analysis showed an 89% correlation between the theoretical and scaled experimental stiffness profiles for values from 0.5 to 2.5. The proposed architecture enables decoupled optimization of impact safety and positional precision, offering a clinically adaptable solution for hemiparetic gait assistance.

MicromachinesVol. 16(12)
University of Shanghai for Science and Technology (CN), Xi’an University of Posts and Telecommunications (CN)
Aarhus Universitet
Openalex Percentile: Top 63%
Prosthetics and Rehabilitation Robotics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Design, Modeling, and Testing—A Compact Variable-Stiffness Actuator for Knee Joint Dimensions — Guoning Si, Zhuo Zhang, et al. · Micromachines (2025) | TGRS Research Map | TGRS