Development and evaluation of the Frankfurt Intubation Device, a user-centred design for enhanced patient safety

Background Endotracheal intubation can be a lifesaving procedure but poses risks, particularly in emergency situations. Inexperienced users benefit from user-friendly devices for successful intubation. Methods This feasibility study assessed the Frankfurt Intubation Device (FID), developed for ease of use with early aspiration protection and a redundant light source. 30 medical students with minimal intubation experience were randomised to start with endotracheal tube (ET)+stylet or FID. After standardised instruction, each performed one manikin intubation per device. Usability was assessed on a 10-point Likert scale (1–10), with lower scores indicating better usability. Results Participants rated the FID as easy to use (median (Md)=3 (2;3)) and suitable for everyday practice (Md=3 (2;4)). They preferred the FID (Md=3 (2;4)) and believed it could become a new standard, requiring minimal training (Md=2 (1.25;3)) while enhancing safety (Md=2 (2;3)). No significant difference in preference was found between FID and ET+stylet. Time to airway protection was 13 (11;16.5) s and 13 (11.5;20) s for ET+stylet, and 16 (13;19.5) s and 14 (12;15.5) s for FID. Time to ventilation increased for ET+stylet from 18 (16.5;23) s to 20 (18;28) s, but decreased for FID from 38 (34.5;46) s to 34 (30.5;38) s. All participants succeeded on the first attempt without dental trauma. Conclusion The FID is easy to use, requires minimal training and shows strong acceptance. The time to ventilation was slower than with ET+stylet in this manikin model but may improve in difficult airways. An industry cooperation is aspired.

Authors

Institutions

Publication Details

Journal
BMJ Innovations
Published
2025-10-01
DOI
https://doi.org/10.1136/bmjinnov-2024-001332
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and evaluation of the Frankfurt Intubation Device, a user-centred design for enhanced patient safety

Benjamin Friedrichson, Kai Zacharowski, Haitham Mutlak, Jan Andreas Kloka et al.
BMJ Innovations
Airway Management and Intubation Techniques
article

Development and evaluation of the Frankfurt Intubation Device, a user-centred design for enhanced patient safety

Benjamin Friedrichson, Kai Zacharowski, Haitham Mutlak, Jan Andreas Kloka, Sebastian Zinn, Hans Magnus Haveraaen, Broder Romeike, Udo Seel
article en

Abstract

Background Endotracheal intubation can be a lifesaving procedure but poses risks, particularly in emergency situations. Inexperienced users benefit from user-friendly devices for successful intubation. Methods This feasibility study assessed the Frankfurt Intubation Device (FID), developed for ease of use with early aspiration protection and a redundant light source. 30 medical students with minimal intubation experience were randomised to start with endotracheal tube (ET)+stylet or FID. After standardised instruction, each performed one manikin intubation per device. Usability was assessed on a 10-point Likert scale (1–10), with lower scores indicating better usability. Results Participants rated the FID as easy to use (median (Md)=3 (2;3)) and suitable for everyday practice (Md=3 (2;4)). They preferred the FID (Md=3 (2;4)) and believed it could become a new standard, requiring minimal training (Md=2 (1.25;3)) while enhancing safety (Md=2 (2;3)). No significant difference in preference was found between FID and ET+stylet. Time to airway protection was 13 (11;16.5) s and 13 (11.5;20) s for ET+stylet, and 16 (13;19.5) s and 14 (12;15.5) s for FID. Time to ventilation increased for ET+stylet from 18 (16.5;23) s to 20 (18;28) s, but decreased for FID from 38 (34.5;46) s to 34 (30.5;38) s. All participants succeeded on the first attempt without dental trauma. Conclusion The FID is easy to use, requires minimal training and shows strong acceptance. The time to ventilation was slower than with ET+stylet in this manikin model but may improve in difficult airways. An industry cooperation is aspired.

BMJ InnovationsVol. 11(4)
Goethe University Frankfurt (DE), University Hospital Frankfurt (DE), Sana Klinikum Offenbach (DE)
Openalex Percentile: Top 44%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.