Impedimetric Graphene-Based Gold Immunosensor for Colorectal Cancer Biomarker CCSP-2 Detection

Electrochemical immunosensors are emerging as promising tools for cancer detection due to their simplicity, portability, and high sensitivity. Colorectal cancer (CRC), the third most common cancer in the United States, remains challenging to diagnose early, as the standard method, colonoscopy, is invasive and often avoided. To address this gap, a graphene-gold-based immunosensor was developed for the early detection of CRC by targeting colon cancer-secreted protein-2 (CCSP-2), a biomarker overexpressed in CRC patients. The sensor was fabricated by attaching graphene oxide (GO) to a gold (Au) electrode using 4-aminothiophenol (4-ATP) as a linker, followed by immobilization of CCSP-2 antibodies (Anti-CCSP-2) and blocking with bovine serum albumin (BSA). Characterization of the immunosensor using linear sweep voltammetry (LSV), Raman spectroscopy, cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS) confirmed successful modifications. The attachment of graphene oxide contributed to an enhancement in current response, likely due to partial reduction and improved electron transfer at the modified surface. The sensor demonstrated a good linear response (<i>R</i> <sup>2</sup> = 0.979) to CCSP-2 antigen (CCSP-2) concentrations ranging from 1 ng/μL to 100 ng/μL, with a limit of detection (LOD) of 0.17 ng/μL and a sensitivity of 0.031 (ng/μL)<sup>-1</sup>. Selectivity was validated using CRC cell extracts (CACO-2) and human kidney extracts (HEK), showing a more significant signal for CACO-2. These findings suggest that the developed immunosensor is a reliable and sensitive platform for CCSP-2 detection, with the potential for adaptation as a point-of-care device for early CRC screening.

Authors

Institutions

Publication Details

Journal
ACS Omega
Published
2025-11-30
DOI
https://doi.org/10.1021/acsomega.5c01338
Primary Topic
Advanced biosensing and bioanalysis techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impedimetric Graphene-Based Gold Immunosensor for Colorectal Cancer Biomarker CCSP-2 Detection

Ramonita Díaz-Ayala, Carlos R. Cabrera, Ruma Paul, Yermary Morales‐Lozada et al.
ACS Omega
Advanced biosensing and bioanalysis techniques
article

Impedimetric Graphene-Based Gold Immunosensor for Colorectal Cancer Biomarker CCSP-2 Detection

Ramonita Díaz-Ayala, Carlos R. Cabrera, Ruma Paul, Yermary Morales‐Lozada, Brian J. Sánchez Colón
article en

Abstract

Electrochemical immunosensors are emerging as promising tools for cancer detection due to their simplicity, portability, and high sensitivity. Colorectal cancer (CRC), the third most common cancer in the United States, remains challenging to diagnose early, as the standard method, colonoscopy, is invasive and often avoided. To address this gap, a graphene-gold-based immunosensor was developed for the early detection of CRC by targeting colon cancer-secreted protein-2 (CCSP-2), a biomarker overexpressed in CRC patients. The sensor was fabricated by attaching graphene oxide (GO) to a gold (Au) electrode using 4-aminothiophenol (4-ATP) as a linker, followed by immobilization of CCSP-2 antibodies (Anti-CCSP-2) and blocking with bovine serum albumin (BSA). Characterization of the immunosensor using linear sweep voltammetry (LSV), Raman spectroscopy, cyclic voltammetry (CV), and electrochemical impedance spectroscopy (EIS) confirmed successful modifications. The attachment of graphene oxide contributed to an enhancement in current response, likely due to partial reduction and improved electron transfer at the modified surface. The sensor demonstrated a good linear response (<i>R</i> <sup>2</sup> = 0.979) to CCSP-2 antigen (CCSP-2) concentrations ranging from 1 ng/μL to 100 ng/μL, with a limit of detection (LOD) of 0.17 ng/μL and a sensitivity of 0.031 (ng/μL)<sup>-1</sup>. Selectivity was validated using CRC cell extracts (CACO-2) and human kidney extracts (HEK), showing a more significant signal for CACO-2. These findings suggest that the developed immunosensor is a reliable and sensitive platform for CCSP-2 detection, with the potential for adaptation as a point-of-care device for early CRC screening.

ACS OmegaVol. 10(48)
The University of Texas at El Paso (US), University of Puerto Rico at Río Piedras (PR)
Openalex Percentile: Top 66%
Advanced biosensing and bioanalysis techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.