HyperSight CBCT image quality and metal artifact reduction for adaptive head and neck radiotherapy: Results from a prospective clinical trial
HyperSight iCBCT MAR significantly reduced artifacts compared to HyperSight iCBCT Acuros, TrueBeam and FBCT, making it particularly beneficial for patients with metal implants. Both HyperSight reconstructions demonstrated improved soft-tissue HU accuracy and contrast compared to TrueBeam, however the iCBCT Acuros algorithm may be preferred when metal-induced artifacts are not a concern. These results support the suitability of HyperSight images in adaptive treatment workflows requiring accurate image quality, even with severe metal artifacts.