RQR beam qualities measurement with the Timepix4 detector

This work presents, to the best of the authors’ knowledge, the first attempt at reconstructing X-ray spectra with a hybrid pixel detector for radiation protection applications in the scattered field around fluoroscopy devices. The performance of the Timepix4 detector was evaluated under reference conditions using RQR beam qualities. A dedicated methodology was developed to correct for detection efficiency, explicitly accounting for contributions from Compton scattering and the photoelectric effect. Accuracy was assessed through Monte Carlo simulations and comparisons with published RQR reference spectra. In addition, particle fluence measurements obtained with Timepix4 were validated against values derived from kerma factors and ionisation chamber measurements performed in a calibration laboratory. The results show excellent agreement between reconstructed and reference spectra up to 70keV, with residuals below 1.7% and relative photon fluence shifts within 5%. At higher energies, discrepancies increase due to the low detection efficiency of the silicon sensor, requiring large correction factors and leading to greater statistical fluctuations. Nevertheless, the system demonstrated acceptable performance up to 120keV, with fluence shifts remaining below 16%. Linearity was confirmed across all beam qualities and intensities investigated, consistent with previous studies on earlier Timepix ASIC generations. Overall, these findings demonstrate that Timepix4 can provide reliable spectral and fluence measurements in reference X-ray beams, offering valuable information for radiation protection studies in scattered fields where conventional detectors are of limited use. Future improvements in readout bandwidth and spectral deconvolution methods are expected to extend its applicability to higher energies and clinical environments. •Spectral measurements were taken in reference conditions with Timepix4.•Results show strong agreement with published reference spectra.•Consistent results in terms of particle fluence were found up to 120 keV.•Timepix4 shows great potential for spectral measurements in interventional radiology.

Authors

Publication Details

Journal
Radiation Measurements
Published
2025-11-28
DOI
https://doi.org/10.1016/j.radmeas.2025.107570
Citations
1
Primary Topic
Radiation Shielding Materials Analysis
Type
article
Field-Weighted Citation Impact
0.52
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

RQR beam qualities measurement with the Timepix4 detector

François Bochud, L. Tlustos, Pierre Carbonez, M. Campbell et al.
1 citations
Radiation Measurements
Radiation Shielding Materials Analysis
0.52
article

RQR beam qualities measurement with the Timepix4 detector

François Bochud, L. Tlustos, Pierre Carbonez, M. Campbell, J. Damet, T. Genetay
article en
1 citations

Abstract

This work presents, to the best of the authors’ knowledge, the first attempt at reconstructing X-ray spectra with a hybrid pixel detector for radiation protection applications in the scattered field around fluoroscopy devices. The performance of the Timepix4 detector was evaluated under reference conditions using RQR beam qualities. A dedicated methodology was developed to correct for detection efficiency, explicitly accounting for contributions from Compton scattering and the photoelectric effect. Accuracy was assessed through Monte Carlo simulations and comparisons with published RQR reference spectra. In addition, particle fluence measurements obtained with Timepix4 were validated against values derived from kerma factors and ionisation chamber measurements performed in a calibration laboratory. The results show excellent agreement between reconstructed and reference spectra up to 70keV, with residuals below 1.7% and relative photon fluence shifts within 5%. At higher energies, discrepancies increase due to the low detection efficiency of the silicon sensor, requiring large correction factors and leading to greater statistical fluctuations. Nevertheless, the system demonstrated acceptable performance up to 120keV, with fluence shifts remaining below 16%. Linearity was confirmed across all beam qualities and intensities investigated, consistent with previous studies on earlier Timepix ASIC generations. Overall, these findings demonstrate that Timepix4 can provide reliable spectral and fluence measurements in reference X-ray beams, offering valuable information for radiation protection studies in scattered fields where conventional detectors are of limited use. Future improvements in readout bandwidth and spectral deconvolution methods are expected to extend its applicability to higher energies and clinical environments. •Spectral measurements were taken in reference conditions with Timepix4.•Results show strong agreement with published reference spectra.•Consistent results in terms of particle fluence were found up to 120 keV.•Timepix4 shows great potential for spectral measurements in interventional radiology.

Radiation MeasurementsVol. 191
Openalex Percentile: Top 30%
Radiation Shielding Materials Analysis
0.52
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.