Repeatability and reproducibility of hemodynamic measurements by intracranial 4D Flow MRI: a multi‑vendor and multi‑software cross‑over comparison study
Scanning protocols by different vendors appear to result in smaller variations in flow rate compared to using different post-processing software methods. It is recommended to employ single post-processing software or conduct software calibration when using multiple software platforms to ensure consistency in multi-center clinical studies.