Subcutaneous Implantable Cardiac Defibrillator Parameters After 65kg Weight Gain – Not What They Were – A Case Report

Abstract Background Subcutaneous implantable cardiac defibrillator (SICD) efficacy can be affected by increased body habitus due to R wave undersensing and unsuccessful defibrillation from ineffectual current delivery. Case Summary Our patient underwent SICD implantation for idiopathic VF. Appropriate shocks were delivered over the next 4 years for episodes of VF storm. However, following this his weight increased from 102kg to 167kg. With this his R wave values in all vectors fell, and device impedance rose to 210Ω. This value was outside manufacturer recommendations and where its efficacy in treating ventricular arrhythmia is unknown. Due to this concern, a full system revision procedure occurred, following which his impedance fell to 80Ω, and he successfully underwent defibrillation testing. At follow up he had remained arrhythmia free with normal device parameters and had commenced a weight loss programme. Discussion Increased body habitus can have profound effects of SICD parameters due to the high resistance of adipose tissue. Parameters and device efficacy can be optimised with device and lead placement adjacent to the chest wall and sternum respectively.

Authors

Institutions

Publication Details

Journal
European Heart Journal - Case Reports
Published
2025-11-27
DOI
https://doi.org/10.1093/ehjcr/ytaf614
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Subcutaneous Implantable Cardiac Defibrillator Parameters After 65kg Weight Gain – Not What They Were – A Case Report

Pier D. Lambiase, J. Boullin, Richard Balasubramaniam, Christopher Monkhouse et al.
European Heart Journal - Case Reports
Cardiac pacing and defibrillation studies
article

Subcutaneous Implantable Cardiac Defibrillator Parameters After 65kg Weight Gain – Not What They Were – A Case Report

Pier D. Lambiase, J. Boullin, Richard Balasubramaniam, Christopher Monkhouse, Alexander P. Bates
article en

Abstract

Abstract Background Subcutaneous implantable cardiac defibrillator (SICD) efficacy can be affected by increased body habitus due to R wave undersensing and unsuccessful defibrillation from ineffectual current delivery. Case Summary Our patient underwent SICD implantation for idiopathic VF. Appropriate shocks were delivered over the next 4 years for episodes of VF storm. However, following this his weight increased from 102kg to 167kg. With this his R wave values in all vectors fell, and device impedance rose to 210Ω. This value was outside manufacturer recommendations and where its efficacy in treating ventricular arrhythmia is unknown. Due to this concern, a full system revision procedure occurred, following which his impedance fell to 80Ω, and he successfully underwent defibrillation testing. At follow up he had remained arrhythmia free with normal device parameters and had commenced a weight loss programme. Discussion Increased body habitus can have profound effects of SICD parameters due to the high resistance of adipose tissue. Parameters and device efficacy can be optimised with device and lead placement adjacent to the chest wall and sternum respectively.

European Heart Journal - Case Reports
St Bartholomew's Hospital (GB), Royal Bournemouth Hospital (GB), University College London (GB)
Openalex Percentile: Top 46%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.