Application of TLD Dosimeters (CaSO₄: Dy and LiF:Mg,Ti) for Environmental Assessment of H*(10) in Field Studies.

The increasing use of ionizing radiation across various sectors of society underscores the need for effective monitoring of occupationally exposed individuals (OEIs), the public, and the environment. Within this framework, environmental radiation protection programs adopt the ambient dose equivalent H*(10) as the operational quantity, as defined by the International Commission on Radiological Protection (ICRP) and recommended by the International Atomic Energy Agency (IAEA), due to its suitability for estimating the risk of external exposure under field conditions. This study assessed the ambient dose equivalent H*(10) through a field survey conducted around the perimeter of the IPEN facilities. Eight monitoring points were selected near thirteen already existing environmental sampling stations, covering approximately 60% of the area. The monitoring was structured in three stages: (i) a comparative analysis between monthly and quarterly measurements using CaSO₄:Dy thermoluminescent detectors (TLDs), which showed proportional responses; (ii) a comparison between quarterly measurements with CaSO₄:Dy and LiF:Mg,Ti TLDs, which demonstrated equivalent results; and (iii) a comparison between the quarterly CaSO₄:Dy data from this study and those obtained in the Environmental Radiological Monitoring Program of the Center for Metrology of Ionizing Radiations of IPEN (ERMP/CEMRI-IPEN), which indicated consistent and satisfactory results, confirming the reliability of using H*(10) in the applied monitoring system. Statistical and uncertainty analyses further confirmed the robustness of the environmental monitoring with thermoluminescent dosimeters. Proper detector selection remains essential to ensure accuracy, minimize variability, and enhance environmental radiation monitoring programs.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2025-11-26
DOI
https://doi.org/10.15392/2319-0612.2025.2925
Primary Topic
Radioactive contamination and transfer
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Application of TLD Dosimeters (CaSO₄: Dy and LiF:Mg,Ti) for Environmental Assessment of H*(10) in Field Studies.

T. Soares, Linda V.E. Caldas, V. P. Campos, L.L. Campos
Brazilian Journal of Radiation Sciences
Radioactive contamination and transfer
article

Application of TLD Dosimeters (CaSO₄: Dy and LiF:Mg,Ti) for Environmental Assessment of H*(10) in Field Studies.

T. Soares, Linda V.E. Caldas, V. P. Campos, L.L. Campos
article en

Abstract

The increasing use of ionizing radiation across various sectors of society underscores the need for effective monitoring of occupationally exposed individuals (OEIs), the public, and the environment. Within this framework, environmental radiation protection programs adopt the ambient dose equivalent H*(10) as the operational quantity, as defined by the International Commission on Radiological Protection (ICRP) and recommended by the International Atomic Energy Agency (IAEA), due to its suitability for estimating the risk of external exposure under field conditions. This study assessed the ambient dose equivalent H*(10) through a field survey conducted around the perimeter of the IPEN facilities. Eight monitoring points were selected near thirteen already existing environmental sampling stations, covering approximately 60% of the area. The monitoring was structured in three stages: (i) a comparative analysis between monthly and quarterly measurements using CaSO₄:Dy thermoluminescent detectors (TLDs), which showed proportional responses; (ii) a comparison between quarterly measurements with CaSO₄:Dy and LiF:Mg,Ti TLDs, which demonstrated equivalent results; and (iii) a comparison between the quarterly CaSO₄:Dy data from this study and those obtained in the Environmental Radiological Monitoring Program of the Center for Metrology of Ionizing Radiations of IPEN (ERMP/CEMRI-IPEN), which indicated consistent and satisfactory results, confirming the reliability of using H*(10) in the applied monitoring system. Statistical and uncertainty analyses further confirmed the robustness of the environmental monitoring with thermoluminescent dosimeters. Proper detector selection remains essential to ensure accuracy, minimize variability, and enhance environmental radiation monitoring programs.

Brazilian Journal of Radiation SciencesVol. 13(3A (Suppl.))
Institut für Psychologische Psychotherapie (DE)
Fundação de Amparo à Pesquisa do Estado de São Paulo, Conselho Nacional de Desenvolvimento Científico e Tecnológico
Openalex Percentile: Top 65%
Radioactive contamination and transfer
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.