A Novel SEM Image Processing Approach for Evaluating Sterilization Effects on Polymeric Medical Devices: Validation Against Traditional EDX Analysis

This study aimed to evaluate the impact of UVC (Ultraviolet C Radiation), detergent foam, and alcohol (70%) sterilization methods on the surface morphology of acrylonitrile-butadiene-styrene (ABS) specimens using a novel SEM (Scanning Electron Microscope) image processing approach. Twelve 3D-printed specimens were prepared, and five concentric circular regions of interest (ROIs) per specimen were analyzed. Three quantitative descriptors-defect area fraction, anisotropy ratio, and RMS (Root Mean Square) roughness-were extracted to assess surface alterations. To validate the image-based findings, EDX (Energy-Dispersive X-ray Spectroscopy) elemental analysis for carbon (C), nitrogen (N), and oxygen (O) was employed as a complementary and traditional benchmark technique. Statistical comparisons and <i>p</i>-value heat maps revealed strong convergence between SEM and EDX results. UVC sterilization consistently preserved surface morphology and elemental stability, showing the lowest defect fraction (<i>p</i> = 0.2684), balanced anisotropy (<i>p</i> = 0.02481), and minimal oxygen incorporation (O = 7.6). Foam sterilization produced intermediate effects, with significant anisotropy changes (<i>p</i> = 0.007456) and reduced nitrogen (19.6). Alcohol sterilization induced the most severe damage, characterized by high defect density, increased roughness, and elemental imbalance (N = 17.3, O = 13.9), confirming oxidative degradation. The convergence of SEM and EDX outcomes demonstrates that SEM image processing is a reliable novel method validated by traditional elemental analysis. Together, these approaches provide a robust framework for ranking sterilization efficacy, with UVC identified as the most favorable method, detergent foam as an acceptable alternative, and alcohol as the least effective due to its destabilizing effects.

Authors

Institutions

Publication Details

Journal
Polymers
Published
2025-11-27
DOI
https://doi.org/10.3390/polym17233156
Primary Topic
Medical Device Sterilization and Disinfection
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Novel SEM Image Processing Approach for Evaluating Sterilization Effects on Polymeric Medical Devices: Validation Against Traditional EDX Analysis

Rashed Almousa, Ibrahim AlMohimeed, Ahmad Alassaf, Talal Alharbi et al.
Polymers
Medical Device Sterilization and Disinfection
article

A Novel SEM Image Processing Approach for Evaluating Sterilization Effects on Polymeric Medical Devices: Validation Against Traditional EDX Analysis

Rashed Almousa, Ibrahim AlMohimeed, Ahmad Alassaf, Talal Alharbi, Vidan F. Ghoneim, Abdulrahman Alduraywish, Mohamed Abdelkader Aboamer
article en

Abstract

This study aimed to evaluate the impact of UVC (Ultraviolet C Radiation), detergent foam, and alcohol (70%) sterilization methods on the surface morphology of acrylonitrile-butadiene-styrene (ABS) specimens using a novel SEM (Scanning Electron Microscope) image processing approach. Twelve 3D-printed specimens were prepared, and five concentric circular regions of interest (ROIs) per specimen were analyzed. Three quantitative descriptors-defect area fraction, anisotropy ratio, and RMS (Root Mean Square) roughness-were extracted to assess surface alterations. To validate the image-based findings, EDX (Energy-Dispersive X-ray Spectroscopy) elemental analysis for carbon (C), nitrogen (N), and oxygen (O) was employed as a complementary and traditional benchmark technique. Statistical comparisons and <i>p</i>-value heat maps revealed strong convergence between SEM and EDX results. UVC sterilization consistently preserved surface morphology and elemental stability, showing the lowest defect fraction (<i>p</i> = 0.2684), balanced anisotropy (<i>p</i> = 0.02481), and minimal oxygen incorporation (O = 7.6). Foam sterilization produced intermediate effects, with significant anisotropy changes (<i>p</i> = 0.007456) and reduced nitrogen (19.6). Alcohol sterilization induced the most severe damage, characterized by high defect density, increased roughness, and elemental imbalance (N = 17.3, O = 13.9), confirming oxidative degradation. The convergence of SEM and EDX outcomes demonstrates that SEM image processing is a reliable novel method validated by traditional elemental analysis. Together, these approaches provide a robust framework for ranking sterilization efficacy, with UVC identified as the most favorable method, detergent foam as an acceptable alternative, and alcohol as the least effective due to its destabilizing effects.

PolymersVol. 17(23)
Princess Nourah bint Abdulrahman University (SA), Qassim University (SA), Majmaah University (SA)
Princess Nourah Bint Abdulrahman University, Majmaah University
Openalex Percentile: Top 65%
Medical Device Sterilization and Disinfection
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.