Monte Carlo simulation to optimize measurement geometry of continuous air monitor with Si semiconductor detector

One of the challenges in real-time monitoring of airborne alpha-emitting radionuclides using Continuous Air Monitors (CAMs) is to discriminate alpha particles emitted by artificial radionuclides from those emitted by natural radionuclides, as the alpha energy spectrum is easily degraded due to its high stopping power. Although the measurement geometry significantly affects the alpha energy spectrum, there are no reports that have investigated the optimized geometry. This study conducted a Monte Carlo simulation to optimize the measurement geometry of the CAMs. The CAM was modelled using a radiation transport simulation code, and the counting efficiencies for 218 Po and 239 Pu and the overlapping ratio were simulated. The Minimum Detectable Concentration (MDC) of 239 Pu was then estimated as an indicator to optimize the geometry. Although a constant counting efficiency is generally used for the same geometry regardless of the alpha energy, the counting efficiency differed up to ∼0.04 by the alpha energy depending on the measurement geometry. The overlapping ratio for a 50-mm diameter detector ranged from 0.22–0.35, while those for a 25-mm diameter detector was below 0.25. This discrepancy is because of the incident angle distribution biased towards large angles. The lowest MDC was estimated to be ∼3 Bq m -3 at 20 L min -1 under 25 Bq m -3 of 218 Po. The MDC was low for small filter diameters and Filter-to-Detector Distances (FDDs) under most of the simulated conditions. The results indicate that a shorter filter diameter and FDD would be preferable for alpha spectrometry under normal air pressure. • Measurement geometry of continuous air monitors was optimized. • Counting efficiency differed by the alpha energy depending on the geometry. • Biased incident angle distribution resulted in high overlapping ratio. • Short filter diameters and filter-to-detector distances would be preferred.

Authors

Institutions

Publication Details

Journal
Radiation Physics and Chemistry
Published
2025-11-27
DOI
https://doi.org/10.1016/j.radphyschem.2025.113507
Primary Topic
Radioactivity and Radon Measurements
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo simulation to optimize measurement geometry of continuous air monitor with Si semiconductor detector

Masahiro Hosoda, Shinji Tokonami, Yasutaka Omori, Kazuki Iwaoka et al.
Radiation Physics and Chemistry
Radioactivity and Radon Measurements
article

Monte Carlo simulation to optimize measurement geometry of continuous air monitor with Si semiconductor detector

Masahiro Hosoda, Shinji Tokonami, Yasutaka Omori, Kazuki Iwaoka, Hiroki Hashimoto, Ryohei Yamada, Yuki Tamakuma, Manaya Taoka
article en

Abstract

One of the challenges in real-time monitoring of airborne alpha-emitting radionuclides using Continuous Air Monitors (CAMs) is to discriminate alpha particles emitted by artificial radionuclides from those emitted by natural radionuclides, as the alpha energy spectrum is easily degraded due to its high stopping power. Although the measurement geometry significantly affects the alpha energy spectrum, there are no reports that have investigated the optimized geometry. This study conducted a Monte Carlo simulation to optimize the measurement geometry of the CAMs. The CAM was modelled using a radiation transport simulation code, and the counting efficiencies for 218 Po and 239 Pu and the overlapping ratio were simulated. The Minimum Detectable Concentration (MDC) of 239 Pu was then estimated as an indicator to optimize the geometry. Although a constant counting efficiency is generally used for the same geometry regardless of the alpha energy, the counting efficiency differed up to ∼0.04 by the alpha energy depending on the measurement geometry. The overlapping ratio for a 50-mm diameter detector ranged from 0.22–0.35, while those for a 25-mm diameter detector was below 0.25. This discrepancy is because of the incident angle distribution biased towards large angles. The lowest MDC was estimated to be ∼3 Bq m -3 at 20 L min -1 under 25 Bq m -3 of 218 Po. The MDC was low for small filter diameters and Filter-to-Detector Distances (FDDs) under most of the simulated conditions. The results indicate that a shorter filter diameter and FDD would be preferable for alpha spectrometry under normal air pressure. • Measurement geometry of continuous air monitors was optimized. • Counting efficiency differed by the alpha energy depending on the geometry. • Biased incident angle distribution resulted in high overlapping ratio. • Short filter diameters and filter-to-detector distances would be preferred.

Radiation Physics and ChemistryVol. 241
Hirosaki University (JP), National Institutes for Quantum Science and Technology (JP), Nagasaki University (JP)
Openalex Percentile: Top 50%
Radioactivity and Radon Measurements
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.