X-ray generators play a crucial role in diagnostic imaging and radiation protection, with standards IEC 61267 and ISO 4037-1 governing the beam commissioning process. However, insufficient detail in measurement techniques and omissions in these standards may introduce errors that affect performance evaluation. This study assesses the performance of a modern X-ray generator during beam commissioning, focusing on voltage accuracy, ripple voltage, and measurement techniques in relation to compliance with these standards. A metrological X-ray system was evaluated using a combination of invasive and non-invasive measurements, employing instruments such as a high-voltage divider, oscilloscope, multimeters, X-ray multimeter, and radiation detectors. Voltage deviation, ripple voltage analysis, and real-time spectrum evaluation were conducted under various conditions. The generator met most of the standard requirements, with voltage deviations consistently below 0.5% across different instruments. The comparison between practical peak voltage and average voltage showed minimal differences, suggesting that average voltage measurements could be broadly applicable. Ripple voltage analysis revealed drawbacks in peak-to-peak versus root-mean-square measurements, emphasizing the importance of measurement bandwidth and the calibration challenges associated with invasive methods. Furthermore, ripple limits based on DC offset assumptions do not adequately reflect the actual impact of ripple on dosimetric quantities. Our results indicate that ripple-averaging effects mitigate these deviations, highlighting the need for further research. These findings suggest potential refinements in ISO 4037 and IEC 61267 regarding measurement techniques, ripple characterization, and uncertainty specifications. While the modern X-ray generator demonstrates solid compliance with current standards, clearer guidelines on voltage accuracy and ripple measurement are necessary. Standard revisions should address measurement bandwidth, discrepancies in ripple measurements, and practical instrument limitations to improve accuracy and ensure reliable X-ray system assessments in clinical and research applications. • Evaluated compliance of a modern X-ray generator within beam-commissioning process. • Critical review of the standards that govern X-ray beam commissioning. • Assessment of different measurement instruments and techniques. • Pinpointed areas for refinement in the relevant standards. • Outlined best practices to improve measurement harmonization.

Authors

Institutions

Publication Details

Journal
Radiation Physics and Chemistry
Published
2025-11-27
DOI
https://doi.org/10.1016/j.radphyschem.2025.113510
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

On measurement standards and techniques in performance evaluation of a modern X-ray generator

Ivica Prlić, Igor Lacković, Luka Pavelić, Gabriela Jazvac et al.
Radiation Physics and Chemistry
Radiation Dose and Imaging
article

On measurement standards and techniques in performance evaluation of a modern X-ray generator

Ivica Prlić, Igor Lacković, Luka Pavelić, Gabriela Jazvac, Ana Marija Kožuljević
article en

Abstract

X-ray generators play a crucial role in diagnostic imaging and radiation protection, with standards IEC 61267 and ISO 4037-1 governing the beam commissioning process. However, insufficient detail in measurement techniques and omissions in these standards may introduce errors that affect performance evaluation. This study assesses the performance of a modern X-ray generator during beam commissioning, focusing on voltage accuracy, ripple voltage, and measurement techniques in relation to compliance with these standards. A metrological X-ray system was evaluated using a combination of invasive and non-invasive measurements, employing instruments such as a high-voltage divider, oscilloscope, multimeters, X-ray multimeter, and radiation detectors. Voltage deviation, ripple voltage analysis, and real-time spectrum evaluation were conducted under various conditions. The generator met most of the standard requirements, with voltage deviations consistently below 0.5% across different instruments. The comparison between practical peak voltage and average voltage showed minimal differences, suggesting that average voltage measurements could be broadly applicable. Ripple voltage analysis revealed drawbacks in peak-to-peak versus root-mean-square measurements, emphasizing the importance of measurement bandwidth and the calibration challenges associated with invasive methods. Furthermore, ripple limits based on DC offset assumptions do not adequately reflect the actual impact of ripple on dosimetric quantities. Our results indicate that ripple-averaging effects mitigate these deviations, highlighting the need for further research. These findings suggest potential refinements in ISO 4037 and IEC 61267 regarding measurement techniques, ripple characterization, and uncertainty specifications. While the modern X-ray generator demonstrates solid compliance with current standards, clearer guidelines on voltage accuracy and ripple measurement are necessary. Standard revisions should address measurement bandwidth, discrepancies in ripple measurements, and practical instrument limitations to improve accuracy and ensure reliable X-ray system assessments in clinical and research applications. • Evaluated compliance of a modern X-ray generator within beam-commissioning process. • Critical review of the standards that govern X-ray beam commissioning. • Assessment of different measurement instruments and techniques. • Pinpointed areas for refinement in the relevant standards. • Outlined best practices to improve measurement harmonization.

Radiation Physics and ChemistryVol. 241
University of Zagreb (HR), Sisters of Charity Hospital (HR), ZeptoMetrix (United States) (US), Institute for Medical Research and Occupational Health (HR)
Openalex Percentile: Top 48%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.