THE USE OF MINI C-ARM MACHINES IN MINOR FOOT AND ANKLE PROCEDURES AND INJECTIONS SHOWS REDUCED COSTS AND CARBON EMISSIONS COMPARED TO TRADITIONAL C-ARM MACHINES IMPROVING ENVIRONMENTAL SUSTAINABILITY

The mini C-arm is a surgeon-operated smaller machine compared to the radiographer-operated traditional C-arm machine. There is limited data for the financial and environmental impact of the use of mini C-arms in minor foot and ankle procedures and injection cases. We did a cost and usage analysis including initial purchasing price and yearly operational expenditures of mini C-arms versus traditional C-arms, along with investigating the carbon emissions of each machine. We also studied aspects of the radiation impact of utilising mini C-arms. A database consisting of 156 minor foot and ankle procedures and injections was created. There were operational cost savings of £8,135.26 for the 156 cases over the course of one year when utilising mini C-arms instead of traditional C-arms, excluding radiation safety course fees. The traditional C-arms created 1.4 times increase in carbon emission output compared to mini C-arms, which demonstrated obvious environmental implications. A diagnostic reference level (DRL) was calculated for the database, producing a 48% lower value compared to the local DRL for foot and ankle imaging. The mini C-arm mean total dose area product (DAP) at 1.63cGy.cm2 was found to be reduced compared to the local DRL at 4cGy.cm2. For the minor foot and ankle procedures and injections of this study, mini C-arms have shown significant financial and environmental benefits compared to traditional C-arms. Radiation exposure benefits of the mini C-arms were also seen in this study. Further larger studies could help confirm our findings and determine best practice for the use of these machines.

Authors

Institutions

Publication Details

Journal
Orthopaedic Proceedings
Published
2025-11-27
DOI
https://doi.org/10.1302/1358-992x.2025.15.001
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

THE USE OF MINI C-ARM MACHINES IN MINOR FOOT AND ANKLE PROCEDURES AND INJECTIONS SHOWS REDUCED COSTS AND CARBON EMISSIONS COMPARED TO TRADITIONAL C-ARM MACHINES IMPROVING ENVIRONMENTAL SUSTAINABILITY

Michelle Tetlow, Efstathios Drampalos, James M. Gilchrist, Alastair Dick et al.
Orthopaedic Proceedings
Radiation Dose and Imaging
article

THE USE OF MINI C-ARM MACHINES IN MINOR FOOT AND ANKLE PROCEDURES AND INJECTIONS SHOWS REDUCED COSTS AND CARBON EMISSIONS COMPARED TO TRADITIONAL C-ARM MACHINES IMPROVING ENVIRONMENTAL SUSTAINABILITY

Michelle Tetlow, Efstathios Drampalos, James M. Gilchrist, Alastair Dick, T. Syed, C. Durick
article en

Abstract

The mini C-arm is a surgeon-operated smaller machine compared to the radiographer-operated traditional C-arm machine. There is limited data for the financial and environmental impact of the use of mini C-arms in minor foot and ankle procedures and injection cases. We did a cost and usage analysis including initial purchasing price and yearly operational expenditures of mini C-arms versus traditional C-arms, along with investigating the carbon emissions of each machine. We also studied aspects of the radiation impact of utilising mini C-arms. A database consisting of 156 minor foot and ankle procedures and injections was created. There were operational cost savings of £8,135.26 for the 156 cases over the course of one year when utilising mini C-arms instead of traditional C-arms, excluding radiation safety course fees. The traditional C-arms created 1.4 times increase in carbon emission output compared to mini C-arms, which demonstrated obvious environmental implications. A diagnostic reference level (DRL) was calculated for the database, producing a 48% lower value compared to the local DRL for foot and ankle imaging. The mini C-arm mean total dose area product (DAP) at 1.63cGy.cm2 was found to be reduced compared to the local DRL at 4cGy.cm2. For the minor foot and ankle procedures and injections of this study, mini C-arms have shown significant financial and environmental benefits compared to traditional C-arms. Radiation exposure benefits of the mini C-arms were also seen in this study. Further larger studies could help confirm our findings and determine best practice for the use of these machines.

Orthopaedic ProceedingsVol. 107-B(SUPP_15)
Forth Valley Royal Hospital (GB), University of Glasgow (GB)
Openalex Percentile: Top 48%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.