A multi-distance shared calibration curve for quantitative analysis of manganese in high-manganese steel based on a portable remote LIBS instrument
This work demonstrates that a multi-distance shared calibration curve can quantify Mn at 2000, 2200, and 2400 mm. This method eliminates the need for separate curves for each distance, enabling reliable quantitative analysis with a unified model.
Authors
- Guanghui Niu (ORCID: https://orcid.org/0000-0003-4522-7826)
- Haodong Wu (ORCID: https://orcid.org/0000-0002-4306-0682)
- Lina Luo (ORCID: https://orcid.org/0000-0001-5864-2506)
- Yixiang Duan (ORCID: https://orcid.org/0000-0002-2333-4586)
- Qingyu Lin (ORCID: https://orcid.org/0000-0001-9631-5880)
- Mengjiao Zhu
- Yue‐Qin Tang (ORCID: https://orcid.org/0000-0001-6872-1099)
- Zixin Zhang (ORCID: https://orcid.org/0000-0002-5775-3041)
- Atif Shahbaz
- Kaiming Sha
Institutions
- Sichuan University (CN)
- Intelligent Health (United Kingdom) (GB)
Publication Details
- Journal
- Journal of Analytical Atomic Spectrometry
- Published
- 2025-11-27
- DOI
- https://doi.org/10.1039/d5ja00431d
- Primary Topic
- Laser-induced spectroscopy and plasma
- Type
- article
- Field-Weighted Citation Impact
- 0.00