Quality and Dosimetric Accuracy of Linac-Based Single-Isocenter Treatment Plans for Four to Eighteen Brain Metastases

<b>Background</b>: Stereotactic radiotherapy (SRT) is a promising treatment option for patients with multiple brain metastases (BMs). Using one isocenter instead of a separate isocenter for each BM can reduce the treatment time. This work compares the calculated dose in the treatment planning system with the measured dose using film dosimetry of single-isocenter multi-target (SIMT) SRT for multiple BM. <b>Methods</b>: Fifty patients with 4 to 18 BMs (median = 6, in total 356 BMs) were treated with a single-isocenter non-coplanar LINAC-based treatment with six VMAT arcs. Treatment was performed using RayStation and Elekta Versa HD with Agility multileaf collimator, including a 6D robotic couch. Patient-specific QA measurements were performed with an in-house developed phantom using three layers of GafChromic EBT3 film. Film measurements were analyzed in DoseLab using global gamma with 3% and 1 mm distance-to-agreement criteria. Additionally, secondary dose calculations in Mobius3D were performed with similar gamma criteria. <b>Results</b>: The mean total Paddick conformity index and gradient index were 0.7 ± 0.10 and 5.2 ± 1.9, respectively. Monitor units used were 6321 ± 2510, and mean irradiation time was 600 ± 90 s. The mean global gamma passing rate for all measured films was 94.5 ± 4.6% with 3% and 1 mm criteria, while that of the dose calculations in Mobius3D was 98.2 ± 1.2% with the same criteria. A dependence of gamma passing rates of film measurements on the total PTV volume was observed, whereas such dependence was minimal for Mobius3D. <b>Conclusions</b>: The results demonstrate good agreement between the TPS, film measurements, and independent dose calculations, supporting the dosimetric accuracy of single-isocenter multi-target SRT for treating multiple BMs.

Authors

Institutions

Publication Details

Journal
Cancers
Published
2025-11-26
DOI
https://doi.org/10.3390/cancers17233776
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Quality and Dosimetric Accuracy of Linac-Based Single-Isocenter Treatment Plans for Four to Eighteen Brain Metastases

Anna L. Petoukhova, Lia Versluis, Masomah Hashimzadah, Jeroen A. Crouzen et al.
Cancers
Advanced Radiotherapy Techniques
article

Quality and Dosimetric Accuracy of Linac-Based Single-Isocenter Treatment Plans for Four to Eighteen Brain Metastases

Anna L. Petoukhova, Lia Versluis, Masomah Hashimzadah, Jeroen A. Crouzen, Stephanie L. C. Bogers, Jaap Zindler, Wilhelmus J. van der Star, Marc de Goede
article en

Abstract

<b>Background</b>: Stereotactic radiotherapy (SRT) is a promising treatment option for patients with multiple brain metastases (BMs). Using one isocenter instead of a separate isocenter for each BM can reduce the treatment time. This work compares the calculated dose in the treatment planning system with the measured dose using film dosimetry of single-isocenter multi-target (SIMT) SRT for multiple BM. <b>Methods</b>: Fifty patients with 4 to 18 BMs (median = 6, in total 356 BMs) were treated with a single-isocenter non-coplanar LINAC-based treatment with six VMAT arcs. Treatment was performed using RayStation and Elekta Versa HD with Agility multileaf collimator, including a 6D robotic couch. Patient-specific QA measurements were performed with an in-house developed phantom using three layers of GafChromic EBT3 film. Film measurements were analyzed in DoseLab using global gamma with 3% and 1 mm distance-to-agreement criteria. Additionally, secondary dose calculations in Mobius3D were performed with similar gamma criteria. <b>Results</b>: The mean total Paddick conformity index and gradient index were 0.7 ± 0.10 and 5.2 ± 1.9, respectively. Monitor units used were 6321 ± 2510, and mean irradiation time was 600 ± 90 s. The mean global gamma passing rate for all measured films was 94.5 ± 4.6% with 3% and 1 mm criteria, while that of the dose calculations in Mobius3D was 98.2 ± 1.2% with the same criteria. A dependence of gamma passing rates of film measurements on the total PTV volume was observed, whereas such dependence was minimal for Mobius3D. <b>Conclusions</b>: The results demonstrate good agreement between the TPS, film measurements, and independent dose calculations, supporting the dosimetric accuracy of single-isocenter multi-target SRT for treating multiple BMs.

CancersVol. 17(23)
Medisch Centrum Haaglanden (NL), Huisman (Netherlands) (NL)
Openalex Percentile: Top 54%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.