Commissioning of the RefleXion X1 Linear Accelerator and Development of a Consensus Reference Beam Model
A systematic method for Reflexion planning and dosimetric commissioning was developed including a standardized set of treatment plans for model validation. Inter-institution differences between measurements and beam models suggest a gold beam reference dataset may be achievable with improved QC at installation, but at this time individual, site-specific beam models are necessary to encompass machine-specific variations.
Authors
- R Teboh Forbang
- Chenyang Shen (ORCID: https://orcid.org/0000-0001-6490-6555)
- A Ndlovu
- Jina Chang (ORCID: https://orcid.org/0000-0003-0579-6020)
- W. Tyler Watkins
- Huixiao Chen
- Bin Han (ORCID: https://orcid.org/0000-0001-8179-5762)
- T. Banks (ORCID: https://orcid.org/0000-0003-4786-6891)
- Nataliya Kovalchuk (ORCID: https://orcid.org/0000-0003-2619-7262)
- Srinath S. Maganti
- Murat Sürücü (ORCID: https://orcid.org/0000-0003-1693-6850)
- Chunhui Han (ORCID: https://orcid.org/0000-0002-3717-8473)
- An Liu (ORCID: https://orcid.org/0000-0002-8903-5028)
- Bin Cai (ORCID: https://orcid.org/0000-0002-5915-3664)
- David J. Carlson (ORCID: https://orcid.org/0000-0001-8677-7707)
- Lichung Ku
- D. Dimitriadou
- Zhe Chen (ORCID: https://orcid.org/0009-0005-4935-8921)
- M. Saiful Huq
- Ray Yeng Yang
- R.J. Lalonde
Institutions
- City Of Hope National Medical Center (US)
- Hackensack University Medical Center (US)
- Salem Hospital (US)
- Yale University (US)
- RefleXion Medical (United States) (US)
- UPMC Hillman Cancer Center (US)
- The University of Texas Southwestern Medical Center (US)
- Stanford University (US)
Publication Details
- Journal
- Medical Physics
- Published
- 2025-11-26
- DOI
- https://doi.org/10.1002/mp.70173
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00