Evaluation of Trueness and Precision in Extraoral 3D Facial Scanning Systems Using a 3D-Printed Head Model: An In Vitro Study

<b>Objective</b>: This in vitro study aimed to evaluate and compare the trueness and precision of four extraoral 3D facial scanning systems using a standardized 3D-printed human head model. <b>Methods</b>: A 3D-printed head model with 16 anatomical landmarks and 17 inter-landmark linear distances was fabricated using a high-resolution 3D printer. Caliper measurements were used as reference standards. The model was scanned 15 times by four systems: a handheld scanner (MetiSmile, Shining 3D, Hangzhou, China), a desktop scanner (RAYFace v2.0, Ray Co., Seongnam, Gyeonggi-do, Republic of Korea), and two mobile applications (Heges and Polycam, iPhone 15, Apple Inc., Cupertino, CA, USA). All digital distances were measured in Blender software. To assess intra-observer reliability, all measurements were repeated twice by the same examiner with a 3-week interval between sessions, and intra-class correlation coefficients were calculated using a two-way mixed-effects, single-measurement, absolute-agreement model (ICC 3,1). Trueness, defined as the absolute deviation from the reference caliper values, was compared across scanners using the Kruskal-Wallis test due to its non-normal distribution. Precision, regional trueness and precision values across the four scanners defined as the standard deviation of repeated scans, was analyzed using One-way ANOVA with Tukey post-hoc comparisons for normally distributed datasets (α = 0.05). Distances were measured digitally in Blender software, and trueness (absolute deviation from reference) and precision (standard deviation of repeated scans) were analyzed using the Kruskal-Wallis test and One-way ANOVA with Tukey post hoc comparisons (α = 0.05). <b>Results</b>: The Polycam application demonstrated the highest trueness (0.49 ± 0.32 mm), followed by MetiSmile (0.51 ± 0.36 mm), RAYFace (0.58 ± 0.39 mm), and Heges (0.73 ± 0.42 mm). The MetiSmile scanner showed the highest precision (0.12 ± 0.07 mm), while RAYFace and Polycam exhibited moderate precision (0.28 ± 0.19 mm and 0.15 ± 0.06 mm, respectively). Vertical measurements tended to be more accurate than horizontal ones, and the lower facial region showed smaller deviations; however, these differences were not statistically significant (<i>p</i> > 0.05). <b>Conclusions</b>: MetiSmile achieved the highest precision and Polycam the highest trueness. Although all systems showed mean deviations < 1 mm, only three demonstrated <0.6 mm accuracy (except for Heges scanner). These results suggest that professional and mobile-based scanners can provide clinically acceptable facial data for educational and preliminary digital workflow applications, though further validation under clinical conditions is required. This study provides quantitative evidence on the accuracy and repeatability of commonly available extraoral 3D facial scanning systems under controlled laboratory conditions. The results indicate that both professional-grade and mobile-based scanners can reproduce facial morphology with clinically acceptable deviations, particularly in flat and stable regions such as the forehead and chin. Although only three systems achieved mean trueness below 0.6 mm, all demonstrated errors within 1 mm, sufficient for diagnostic visualization, digital smile design, and preliminary virtual patient modeling. These findings support the safe and cost-effective adoption of extraoral facial scanning in dental education and treatment planning, while emphasizing the need for further validation in real clinical environments where motion, lighting, and soft-tissue variability may affect accuracy.

Authors

Institutions

Publication Details

Journal
Journal of Clinical Medicine
Published
2025-11-26
DOI
https://doi.org/10.3390/jcm14238384
Citations
1
Primary Topic
Orthodontics and Dentofacial Orthopedics
Type
article
Field-Weighted Citation Impact
1.68
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of Trueness and Precision in Extraoral 3D Facial Scanning Systems Using a 3D-Printed Head Model: An In Vitro Study

Hoang Viet, Anh N. Le, Trang Nhat Uyen Doan, Tran Hung Lam et al.
1 citations
Journal of Clinical Medicine
Orthodontics and Dentofacial Orthopedics
1.68
article

Evaluation of Trueness and Precision in Extraoral 3D Facial Scanning Systems Using a 3D-Printed Head Model: An In Vitro Study

Hoang Viet, Anh N. Le, Trang Nhat Uyen Doan, Tran Hung Lam, Phuc Ngoc Nguyen, Khanh Linh Vu, Tue H. Nguyen, Khang Chi Duong
article en
1 citations

Abstract

<b>Objective</b>: This in vitro study aimed to evaluate and compare the trueness and precision of four extraoral 3D facial scanning systems using a standardized 3D-printed human head model. <b>Methods</b>: A 3D-printed head model with 16 anatomical landmarks and 17 inter-landmark linear distances was fabricated using a high-resolution 3D printer. Caliper measurements were used as reference standards. The model was scanned 15 times by four systems: a handheld scanner (MetiSmile, Shining 3D, Hangzhou, China), a desktop scanner (RAYFace v2.0, Ray Co., Seongnam, Gyeonggi-do, Republic of Korea), and two mobile applications (Heges and Polycam, iPhone 15, Apple Inc., Cupertino, CA, USA). All digital distances were measured in Blender software. To assess intra-observer reliability, all measurements were repeated twice by the same examiner with a 3-week interval between sessions, and intra-class correlation coefficients were calculated using a two-way mixed-effects, single-measurement, absolute-agreement model (ICC 3,1). Trueness, defined as the absolute deviation from the reference caliper values, was compared across scanners using the Kruskal-Wallis test due to its non-normal distribution. Precision, regional trueness and precision values across the four scanners defined as the standard deviation of repeated scans, was analyzed using One-way ANOVA with Tukey post-hoc comparisons for normally distributed datasets (α = 0.05). Distances were measured digitally in Blender software, and trueness (absolute deviation from reference) and precision (standard deviation of repeated scans) were analyzed using the Kruskal-Wallis test and One-way ANOVA with Tukey post hoc comparisons (α = 0.05). <b>Results</b>: The Polycam application demonstrated the highest trueness (0.49 ± 0.32 mm), followed by MetiSmile (0.51 ± 0.36 mm), RAYFace (0.58 ± 0.39 mm), and Heges (0.73 ± 0.42 mm). The MetiSmile scanner showed the highest precision (0.12 ± 0.07 mm), while RAYFace and Polycam exhibited moderate precision (0.28 ± 0.19 mm and 0.15 ± 0.06 mm, respectively). Vertical measurements tended to be more accurate than horizontal ones, and the lower facial region showed smaller deviations; however, these differences were not statistically significant (<i>p</i> > 0.05). <b>Conclusions</b>: MetiSmile achieved the highest precision and Polycam the highest trueness. Although all systems showed mean deviations < 1 mm, only three demonstrated <0.6 mm accuracy (except for Heges scanner). These results suggest that professional and mobile-based scanners can provide clinically acceptable facial data for educational and preliminary digital workflow applications, though further validation under clinical conditions is required. This study provides quantitative evidence on the accuracy and repeatability of commonly available extraoral 3D facial scanning systems under controlled laboratory conditions. The results indicate that both professional-grade and mobile-based scanners can reproduce facial morphology with clinically acceptable deviations, particularly in flat and stable regions such as the forehead and chin. Although only three systems achieved mean trueness below 0.6 mm, all demonstrated errors within 1 mm, sufficient for diagnostic visualization, digital smile design, and preliminary virtual patient modeling. These findings support the safe and cost-effective adoption of extraoral facial scanning in dental education and treatment planning, while emphasizing the need for further validation in real clinical environments where motion, lighting, and soft-tissue variability may affect accuracy.

Journal of Clinical MedicineVol. 14(23)
Van Lang University (VN)
Openalex Percentile: Top 13%
Orthodontics and Dentofacial Orthopedics
1.68
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.