Computer-assisted vacuum thrombectomy using lightning 12 and lightning flash 16 in acute proximal lower extremity deep vein thrombosis

ObjectiveTo report a single-center experience using the Lightning 12 and Lightning Flash 16 computer-assisted vacuum thrombectomy (CAVT) Indigo System devices for acute proximal lower extremity deep vein thrombosis (DVT).MethodsA retrospective review was conducted of all consecutive patients treated with CAVT for acute (<14 days) proximal lower extremity DVT between January 2023 and September 2024 at academic single center. The primary outcome was technical success, defined as >70% thrombus removal on final completion venography, assessed by at least two different operators.ResultsTwenty-four patients (median age 55.5 years; 58.3% female) with 29 affected limbs were included. Classical May-Thurner syndrome was present in 37.5%. Thrombosis involved the left iliofemoral segment in 44.8%, right iliofemoral in 31.0%, isolated common iliac vein in 6.8%, and inferior vena cava extension in 17.2%. Devices used included Lightning 12 (25.0%) and Lightning Flash 16 (75.0%). Iliac vein stents were placed in 82.8% of limbs. Median operative time was 120 min, blood loss 750 cc, and hospital stay 2 days. No intraoperative complications occurred. Two non-procedure-related deaths occurred within 30 days. Median follow-up was 9 months; three additional late deaths were unrelated to the procedure. The median Villalta score was 1, and VCSS was 1. No significant outcome differences were observed between the two device types.ConclusionThe CAVT Indigo System appears to be a safe and effective treatment option for acute proximal lower extremity DVT, demonstrating high technical success and low complication rates.

Authors

Institutions

Publication Details

Journal
Phlebology The Journal of Venous Disease
Published
2025-11-26
DOI
https://doi.org/10.1177/02683555251404147
Primary Topic
Venous Thromboembolism Diagnosis and Management
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Computer-assisted vacuum thrombectomy using lightning 12 and lightning flash 16 in acute proximal lower extremity deep vein thrombosis

Michel Bergoeing, Mauricio Gonzalez‐Urquijo, Leopoldo A. Mariné, Francisco Valdés et al.
Phlebology The Journal of Venous Disease
Venous Thromboembolism Diagnosis and Management
article

Computer-assisted vacuum thrombectomy using lightning 12 and lightning flash 16 in acute proximal lower extremity deep vein thrombosis

Michel Bergoeing, Mauricio Gonzalez‐Urquijo, Leopoldo A. Mariné, Francisco Valdés, Francisco Vargas, Sebastián Sepúlveda
article en

Abstract

ObjectiveTo report a single-center experience using the Lightning 12 and Lightning Flash 16 computer-assisted vacuum thrombectomy (CAVT) Indigo System devices for acute proximal lower extremity deep vein thrombosis (DVT).MethodsA retrospective review was conducted of all consecutive patients treated with CAVT for acute (<14 days) proximal lower extremity DVT between January 2023 and September 2024 at academic single center. The primary outcome was technical success, defined as >70% thrombus removal on final completion venography, assessed by at least two different operators.ResultsTwenty-four patients (median age 55.5 years; 58.3% female) with 29 affected limbs were included. Classical May-Thurner syndrome was present in 37.5%. Thrombosis involved the left iliofemoral segment in 44.8%, right iliofemoral in 31.0%, isolated common iliac vein in 6.8%, and inferior vena cava extension in 17.2%. Devices used included Lightning 12 (25.0%) and Lightning Flash 16 (75.0%). Iliac vein stents were placed in 82.8% of limbs. Median operative time was 120 min, blood loss 750 cc, and hospital stay 2 days. No intraoperative complications occurred. Two non-procedure-related deaths occurred within 30 days. Median follow-up was 9 months; three additional late deaths were unrelated to the procedure. The median Villalta score was 1, and VCSS was 1. No significant outcome differences were observed between the two device types.ConclusionThe CAVT Indigo System appears to be a safe and effective treatment option for acute proximal lower extremity DVT, demonstrating high technical success and low complication rates.

Phlebology The Journal of Venous Disease
Pontificia Universidad Católica de Chile (CL), Tecnológico de Monterrey (MX)
Openalex Percentile: Top 42%
Venous Thromboembolism Diagnosis and Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.