Simulation and validation of spinal construct testing based on ASTM F1717
Validated FEA reproduced ASTM F1717 construct behavior within prespecified acceptance bounds and can complement early design screening. The parametric analysis highlights diameter as the dominant geometric factor for construct level force response under the tested conditions, whereas length had a comparatively minor influence. These findings indicate that the evaluated screw configurations can enhance structural durability and surgical safety, and they support the use of validated FEA as a reliable alternative to selected physical tests during early stages of implant design.