Comparison of Quantitative Cerebrospinal Fluid Cytology Using Manual Counting and the Sysmex XN–1000

Abstract Background Cerebrospinal fluid (CSF) cytology is essential for diagnosing neurological disorders. Automated analysers offer faster results, but their accuracy in cell differentiation remains debated. This study compares CSF cytology using the Sysmex XN–1000 analyser with manual counting in the Fuchs–Rosenthal (F–R) chamber. Methods A total of 89 CSF samples from patients aged 10 days to 80 years were analysed. Manual counting was performed using the F–R chamber, while the Sysmex XN–1000 measured cell counts in body fluid mode. Statistical comparisons in Medcalc software included the Shapiro–Wilk test, Wilcoxon test, Bland–Altman plots, Mann–Whitney test, and Cohen’s kappa coefficient. Results There were no statistically significant differences between the F–R chamber compared to Sysmex in mononuclear cells (MNc). Statistical significance was defined as P < 0.05. Data presented as median [min–max]): 10 [1–724]/µl vs. 9 [0–670]/µl resp. ( P = 0.1286); and in red blood cells (RBC): 36 [0–18,091]/µl – vs. 0 [0–34,000]/µl resp. ( P = 0.5412). However, there were statistically significant changes in polymorphonuclear cells (PMNc) ( P = 0.0002) using the F–R chamber – 0 [0–1,428]/µl in comparison with Sysmex – 1 [0–1,008]/µl. Cohen’s kappa analysis demonstrated a very good strength agreement ( K = 0.89) in the classification of oligocytosis and pleocytosis. Conclusions The Sysmex XN–1000 is a fast and efficient tool for CSF cytology but it has limitations, especially in differentiating PMNc in siderophage–rich neonatal samples. Although it improves workflow, manual verification remains crucial for diagnostic accuracy.

Authors

Institutions

Publication Details

Journal
Bratislavské lekárske listy/Bratislava medical journal
Published
2025-11-25
DOI
https://doi.org/10.1007/s44411-025-00425-w
Primary Topic
Multiple Sclerosis Research Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Quantitative Cerebrospinal Fluid Cytology Using Manual Counting and the Sysmex XN–1000

Jaroslav Ráček, Daniel Rajdl, J Ženková, Klára Koldušková et al.
Bratislavské lekárske listy/Bratislava medical journal
Multiple Sclerosis Research Studies
article

Comparison of Quantitative Cerebrospinal Fluid Cytology Using Manual Counting and the Sysmex XN–1000

Jaroslav Ráček, Daniel Rajdl, J Ženková, Klára Koldušková, Simona Kukrálová, Pavel Brož, Jana Černá, Eliška Naušová
article en

Abstract

Abstract Background Cerebrospinal fluid (CSF) cytology is essential for diagnosing neurological disorders. Automated analysers offer faster results, but their accuracy in cell differentiation remains debated. This study compares CSF cytology using the Sysmex XN–1000 analyser with manual counting in the Fuchs–Rosenthal (F–R) chamber. Methods A total of 89 CSF samples from patients aged 10 days to 80 years were analysed. Manual counting was performed using the F–R chamber, while the Sysmex XN–1000 measured cell counts in body fluid mode. Statistical comparisons in Medcalc software included the Shapiro–Wilk test, Wilcoxon test, Bland–Altman plots, Mann–Whitney test, and Cohen’s kappa coefficient. Results There were no statistically significant differences between the F–R chamber compared to Sysmex in mononuclear cells (MNc). Statistical significance was defined as P < 0.05. Data presented as median [min–max]): 10 [1–724]/µl vs. 9 [0–670]/µl resp. ( P = 0.1286); and in red blood cells (RBC): 36 [0–18,091]/µl – vs. 0 [0–34,000]/µl resp. ( P = 0.5412). However, there were statistically significant changes in polymorphonuclear cells (PMNc) ( P = 0.0002) using the F–R chamber – 0 [0–1,428]/µl in comparison with Sysmex – 1 [0–1,008]/µl. Cohen’s kappa analysis demonstrated a very good strength agreement ( K = 0.89) in the classification of oligocytosis and pleocytosis. Conclusions The Sysmex XN–1000 is a fast and efficient tool for CSF cytology but it has limitations, especially in differentiating PMNc in siderophage–rich neonatal samples. Although it improves workflow, manual verification remains crucial for diagnostic accuracy.

Bratislavské lekárske listy/Bratislava medical journalVol. 127(1)
Charles University (CZ), University Hospital Plzen (CZ)
Openalex Percentile: Top 50%
Multiple Sclerosis Research Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.