Twiddler-, Reel-, and Ratchet-syndrome: an alternative approach to address the underlying mechanism of complication in cardiac implantable electronic devices

The circumstances leading to the origins of "Twiddler", "Reel", and "Ratchet" in patients with cardiac implantable electronic devices (CIED) have not yet been fully determined. This retrospective study was performed to address different mechanisms that lead to impairment of implanted leads within the context of "Twiddler", "Reel", and "Ratchet". From 2004 and 2023, 5966 patients underwent CIED procedures at our institution. 31 patients with lead impairment by rotation were identified with evidence of pulse generator mobility, lead dislodgement with retraction and / or associated coiling of the lead(s) radiographically or intraoperatively. A review of the literature from 1963 to 2024 identified 216 cases from 165 publications. In less than half of our patients lead retraction could be attested, and a minority of patients had clinical symptoms attributable to lead impairment by rotation. Psychiatric comorbidities and patients´ manipulation could not be ascertained. Thorough follow-up of all patients with CIEDs appears mandatory. Precise implantation techniques from the outset may limit lead impairment by rotation.

Authors

Institutions

Publication Details

Journal
Current Problems in Cardiology
Published
2025-11-24
DOI
https://doi.org/10.1016/j.cpcardiol.2025.103214
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Twiddler-, Reel-, and Ratchet-syndrome: an alternative approach to address the underlying mechanism of complication in cardiac implantable electronic devices

Sigrid Wiesner, Simon Schopka, Andrea Stadlbauer, Christian Stadlbauer et al.
Current Problems in Cardiology
Cardiac pacing and defibrillation studies
article

Twiddler-, Reel-, and Ratchet-syndrome: an alternative approach to address the underlying mechanism of complication in cardiac implantable electronic devices

Sigrid Wiesner, Simon Schopka, Andrea Stadlbauer, Christian Stadlbauer, Ekrem Üçer, Christoph Schmid, Andreas Keyser, Carsten Jungbauer
article en

Abstract

The circumstances leading to the origins of "Twiddler", "Reel", and "Ratchet" in patients with cardiac implantable electronic devices (CIED) have not yet been fully determined. This retrospective study was performed to address different mechanisms that lead to impairment of implanted leads within the context of "Twiddler", "Reel", and "Ratchet". From 2004 and 2023, 5966 patients underwent CIED procedures at our institution. 31 patients with lead impairment by rotation were identified with evidence of pulse generator mobility, lead dislodgement with retraction and / or associated coiling of the lead(s) radiographically or intraoperatively. A review of the literature from 1963 to 2024 identified 216 cases from 165 publications. In less than half of our patients lead retraction could be attested, and a minority of patients had clinical symptoms attributable to lead impairment by rotation. Psychiatric comorbidities and patients´ manipulation could not be ascertained. Thorough follow-up of all patients with CIEDs appears mandatory. Precise implantation techniques from the outset may limit lead impairment by rotation.

Current Problems in CardiologyVol. 51(2)
University Hospital Regensburg (DE), University of Regensburg (DE)
Openalex Percentile: Top 47%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.