<i>Objective.</i>Calf blood pressure (CBP) plays an important role in various clinical applications, such as determining the appropriate cuff pressure for compression therapy and diagnosis of lower extremity vascular diseases, which necessitates the integration of built-in measurement methods within the device. This study aimed to investigate the differences in human CBP assessments resulting from the application of various<i>in vivo</i>external compression strategies.<i>Approach.</i>An experimental procedure incorporating different compression strategies, specially the low-pressure-sustained mode, the slow-inflation (SI) mode and the slow-deflation (SD) mode, was conducted to capture the dynamic responses of the photoplethysmography (PPG) signal for CBP evaluation. Nineteen subjects, including 13 males and 6 females, participated in this experimental study. Feature points related to CBP were extracted from dynamic responses of the PPG signal and subjected to statistical analysis. A lumped parameter model of the human lower extremity was developed to assist in analyzing the biomechanical factors underlying these differences.<i>Main results.</i>The experimental results indicated that the dynamic behaviors of the PPG signal followed clear and consistent patterns, and the CBP value derived from the PPG signal in the SI mode was significantly higher than that obtained in the SD mode. Model simulation results showed that the differences in the evaluated values of the CBP between the SI and SD modes were caused by the different collapse processes of calf arteries and veins.<i>Significance.</i>This study could help understand the forced collapse process of blood vessels in the calf and inspire new ideas on CBP evaluation and personalized compression therapy.

Authors

Institutions

Publication Details

Journal
Physiological Measurement
Published
2025-11-25
DOI
https://doi.org/10.1088/1361-6579/ae241d
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In vivo evaluation of human calf blood pressure by using different external compression strategies

Guifu Wu, Bitian Wang, Shuai Tian, Hanhao Liu et al.
Physiological Measurement
Non-Invasive Vital Sign Monitoring
article

In vivo evaluation of human calf blood pressure by using different external compression strategies

Guifu Wu, Bitian Wang, Shuai Tian, Hanhao Liu, Xuanhao Xu, Yawei Wang, Ruya Li, Yubo Fan
article en

Abstract

<i>Objective.</i>Calf blood pressure (CBP) plays an important role in various clinical applications, such as determining the appropriate cuff pressure for compression therapy and diagnosis of lower extremity vascular diseases, which necessitates the integration of built-in measurement methods within the device. This study aimed to investigate the differences in human CBP assessments resulting from the application of various<i>in vivo</i>external compression strategies.<i>Approach.</i>An experimental procedure incorporating different compression strategies, specially the low-pressure-sustained mode, the slow-inflation (SI) mode and the slow-deflation (SD) mode, was conducted to capture the dynamic responses of the photoplethysmography (PPG) signal for CBP evaluation. Nineteen subjects, including 13 males and 6 females, participated in this experimental study. Feature points related to CBP were extracted from dynamic responses of the PPG signal and subjected to statistical analysis. A lumped parameter model of the human lower extremity was developed to assist in analyzing the biomechanical factors underlying these differences.<i>Main results.</i>The experimental results indicated that the dynamic behaviors of the PPG signal followed clear and consistent patterns, and the CBP value derived from the PPG signal in the SI mode was significantly higher than that obtained in the SD mode. Model simulation results showed that the differences in the evaluated values of the CBP between the SI and SD modes were caused by the different collapse processes of calf arteries and veins.<i>Significance.</i>This study could help understand the forced collapse process of blood vessels in the calf and inspire new ideas on CBP evaluation and personalized compression therapy.

Physiological MeasurementVol. 47(2)
Sun Yat-sen University (CN), Eighth Affiliated Hospital of Sun Yat-sen University, Beihang University (CN)
Openalex Percentile: Top 65%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.