Evaluation of skin pigmentation effect on photoplethysmography signals using a vascular finger phantom with tunable optical and mechanical properties
These findings suggest there is a measurable impact of skin pigmentation on the PPG signal and highlight the need for further research to improve the equity of light-based sensing technologies across all populations. We provide an advancement for future work in developing <i>in vitro</i> models to assess optical sensing performance across diverse skin tones.