Fully automated detection and identification of CSF shunt valves using YOLOv8 and a class-based reference image assignment as a safety mechanism

The study aimed to develop and evaluate an algorithm based on the YOLOv8x framework to automatically detect and identify cerebrospinal fluid (CSF) shunt valves. This approach seeks to streamline the diagnostic process identifying shunt valve types and pressure levels. A retrospective cohort of 2701 anonymized radiographs comprising six types of CSF shunt valves was used. Data augmentation techniques such as flipping, scaling, and mosaic augmentation were applied during training to enhance robustness. The dataset was split into 80% training and 20% testing subsets as part of a 5-fold cross-validation. Validation was conducted on a separate test set of 295 images using metrics such as mean Average Precision (mAP) at intersection over union thresholds of 50% (mAP50) as well as precision, recall, and F1-scores as metrics. Additionally, a class-based reference image assignment system was used to link the detected valves with the corresponding manufacturer images. These paired images were then independently reviewed by two radiologists to assess the accuracy of the algorithm's classifications. The algorithm achieved a weighted mAP50 of 0.884 and a weighted average F1-score of 94.8%. High F1-scores were observed for Codman Certas (99.6%) and Codman Hakim (99.6%), with lower scores for less common valves like proGAV (30.8%). Radiologists were able to identify both correct and incorrect classifications made by the algorithm with 100% accuracy, due to the integrated safety mechanism. This safety mechanism relies on the fully automated linking of detected valves with the corresponding manufacturer images. In Conclusion the automated system demonstrated high efficiency in detecting and classifying CSF shunt valves, significantly simplifying the diagnostic workflow. Moreover, the integration of a robust safety mechanism ensures that potential misclassifications are identified and corrected.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-11-21
DOI
https://doi.org/10.1038/s41598-025-29201-0
Primary Topic
Cerebrospinal fluid and hydrocephalus
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Fully automated detection and identification of CSF shunt valves using YOLOv8 and a class-based reference image assignment as a safety mechanism

Luca Salhöfer, René Hosch, Johannes Haubold, Cornelius Deuschl et al.
Scientific Reports
Cerebrospinal fluid and hydrocephalus
article

Fully automated detection and identification of CSF shunt valves using YOLOv8 and a class-based reference image assignment as a safety mechanism

Luca Salhöfer, René Hosch, Johannes Haubold, Cornelius Deuschl, Lale Umutlu, Hanna Styczen, Sebastian Zensen, Jannis Straus, Michael Forsting, Mathias Meetschen, Felix Nensa, Yan Li, Maharani Budi Santoso
article en

Abstract

The study aimed to develop and evaluate an algorithm based on the YOLOv8x framework to automatically detect and identify cerebrospinal fluid (CSF) shunt valves. This approach seeks to streamline the diagnostic process identifying shunt valve types and pressure levels. A retrospective cohort of 2701 anonymized radiographs comprising six types of CSF shunt valves was used. Data augmentation techniques such as flipping, scaling, and mosaic augmentation were applied during training to enhance robustness. The dataset was split into 80% training and 20% testing subsets as part of a 5-fold cross-validation. Validation was conducted on a separate test set of 295 images using metrics such as mean Average Precision (mAP) at intersection over union thresholds of 50% (mAP50) as well as precision, recall, and F1-scores as metrics. Additionally, a class-based reference image assignment system was used to link the detected valves with the corresponding manufacturer images. These paired images were then independently reviewed by two radiologists to assess the accuracy of the algorithm's classifications. The algorithm achieved a weighted mAP50 of 0.884 and a weighted average F1-score of 94.8%. High F1-scores were observed for Codman Certas (99.6%) and Codman Hakim (99.6%), with lower scores for less common valves like proGAV (30.8%). Radiologists were able to identify both correct and incorrect classifications made by the algorithm with 100% accuracy, due to the integrated safety mechanism. This safety mechanism relies on the fully automated linking of detected valves with the corresponding manufacturer images. In Conclusion the automated system demonstrated high efficiency in detecting and classifying CSF shunt valves, significantly simplifying the diagnostic workflow. Moreover, the integration of a robust safety mechanism ensures that potential misclassifications are identified and corrected.

Scientific ReportsVol. 15(1)
Essen University Hospital (DE)
Openalex Percentile: Top 61%
Cerebrospinal fluid and hydrocephalus
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.