Xact625i vs. PX-375: a comparative study of online XRF ambient multi-metal monitors vs. ICP-MS

Abstract. This study provides a comprehensive evaluation of the Xact625i and PX-375 online energy dispersive X-ray fluorescence (ED-XRF) instruments for real-time trace elemental monitoring at a rural background site over six months. This represents the first direct comparison between the two instruments, assessing their performance against inductively coupled plasma mass spectrometry (ICP-MS), the reference method for elemental analysis. Both instruments demonstrated strong measurement capabilities, with the Xact625i achieving higher sensitivity for trace metals and showing closer agreement with ICP-MS using 24 h averaging of 2 h data (r2=0.89 vs. 0.78 for the PX-375). The PX-375 reported higher overall concentrations, primarily due to overestimations of Si and S. Both instruments correlated well for Ca, Fe, Zn, and Pb, while detection limits affected Ni and Cd. The Xact625i exhibited superior performance in measuring elements such as S, V, and Mn. While correlations with ICP-MS were high, systematic over- and underestimations in absolute concentrations were found, particularly for the PX-375. When directly comparing the two online instruments using raw 2 h data, a strong agreement was observed (mean r2=0.95). However, systematic slope discrepancies persisted, in line with comparisons against ICP-MS. The findings confirm the reliability of these two ED-XRF instruments for high-time-resolution elemental monitoring as a complementary technique to traditional filter analysis, enabling detailed source apportionment studies, improved trend analysis, and more responsive air quality management strategies. Future work comparing ED-XRF to laboratory-based methods could refine harmonisation efforts and address systematic differences in absolute concentrations.

Authors

Institutions

Publication Details

Journal
Atmospheric measurement techniques
Published
2025-11-25
DOI
https://doi.org/10.5194/amt-18-7021-2025
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Xact625i vs. PX-375: a comparative study of online XRF ambient multi-metal monitors vs. ICP-MS

Petra Pokorná, Andrê S. H. Prévôt, Saliou Mbengue, Jakub Ondráček et al.
Atmospheric measurement techniques
X-ray Spectroscopy and Fluorescence Analysis
article

Xact625i vs. PX-375: a comparative study of online XRF ambient multi-metal monitors vs. ICP-MS

Petra Pokorná, Andrê S. H. Prévôt, Saliou Mbengue, Jakub Ondráček, Manousos Ioannis Manousakas, Laurence C. Windell, Jaroslav Schwarz, Vladimir Ždı́mal, Stefanos Papagiannis, Roman Prokeš
article en

Abstract

Abstract. This study provides a comprehensive evaluation of the Xact625i and PX-375 online energy dispersive X-ray fluorescence (ED-XRF) instruments for real-time trace elemental monitoring at a rural background site over six months. This represents the first direct comparison between the two instruments, assessing their performance against inductively coupled plasma mass spectrometry (ICP-MS), the reference method for elemental analysis. Both instruments demonstrated strong measurement capabilities, with the Xact625i achieving higher sensitivity for trace metals and showing closer agreement with ICP-MS using 24 h averaging of 2 h data (r2=0.89 vs. 0.78 for the PX-375). The PX-375 reported higher overall concentrations, primarily due to overestimations of Si and S. Both instruments correlated well for Ca, Fe, Zn, and Pb, while detection limits affected Ni and Cd. The Xact625i exhibited superior performance in measuring elements such as S, V, and Mn. While correlations with ICP-MS were high, systematic over- and underestimations in absolute concentrations were found, particularly for the PX-375. When directly comparing the two online instruments using raw 2 h data, a strong agreement was observed (mean r2=0.95). However, systematic slope discrepancies persisted, in line with comparisons against ICP-MS. The findings confirm the reliability of these two ED-XRF instruments for high-time-resolution elemental monitoring as a complementary technique to traditional filter analysis, enabling detailed source apportionment studies, improved trend analysis, and more responsive air quality management strategies. Future work comparing ED-XRF to laboratory-based methods could refine harmonisation efforts and address systematic differences in absolute concentrations.

Atmospheric measurement techniquesVol. 18(22)
National Centre of Scientific Research "Demokritos" (GR), Masaryk University (CZ), Paul Scherrer Institute (CH), Czech Academy of Sciences, Global Change Research Institute (CZ), Czech Academy of Sciences, Institute of Chemical Process Fundamentals (CZ)
Grantová Agentura České Republiky, HORIZON EUROPE Framework Programme
Openalex Percentile: Top 54%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.