Impact of adjustable resistance clamps on the in-vitro hemolysis assessment of blood pumps

Adjustable resistance clamps are essential to regulate flow and pressure in in-vitro test loops designed to evaluate hemolysis in blood pumps. However, their contribution to overall hemolysis remains unclear. This study investigated the hemolytic potential of adjustable resistance clamps across different resistance settings. Utilizing a single-pass test loop with heparinized bovine blood, the effect of resistance clamps on hemolysis was evaluated at three resistance settings: high (1.5 L/min, 90 mmHg), medium (5 L/min, 60 mmHg), and low (7.5 L/min, 30 mmHg). Two resistance clamps (sudden vs. gradual constriction) were compared at the medium resistance setting. Hemolysis was quantified by measuring plasma free hemoglobin (pfHb) and normalized index of hemolysis (NIH). Furthermore, computational fluid dynamics simulations were used to predict the hemolysis metric for each resistance setting with 3D-recreated replicas. Compared to hemolysis metrics reported for state-of-the-art blood pumps, resistance clamps may affect the hemolysis substantially, with up to 84% of the NIH at medium resistance settings and < 47% at low/high resistance settings, regardless of the resistance shape/profile. Numerically predicted hemolysis metric strongly correlated with experimental results for three resistance settings (r > 0.998). The findings underscore the importance of considering the influence of adjustable resistance clamps while investigating the hemolytic potential of blood pumps.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2025-11-24
DOI
https://doi.org/10.1038/s41598-025-25380-y
Citations
1
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.67

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of adjustable resistance clamps on the in-vitro hemolysis assessment of blood pumps

Martin Stoiber, Daniel Zimpfer, Krishnaraj Narayanaswamy, Laura C. Schmitt et al.
1 citations
Scientific Reports
Mechanical Circulatory Support Devices
0.67
article

Impact of adjustable resistance clamps on the in-vitro hemolysis assessment of blood pumps

Martin Stoiber, Daniel Zimpfer, Krishnaraj Narayanaswamy, Laura C. Schmitt, Marcus Granegger, Michael Röhrich, Xiangyu He
article en
1 citations

Abstract

Adjustable resistance clamps are essential to regulate flow and pressure in in-vitro test loops designed to evaluate hemolysis in blood pumps. However, their contribution to overall hemolysis remains unclear. This study investigated the hemolytic potential of adjustable resistance clamps across different resistance settings. Utilizing a single-pass test loop with heparinized bovine blood, the effect of resistance clamps on hemolysis was evaluated at three resistance settings: high (1.5 L/min, 90 mmHg), medium (5 L/min, 60 mmHg), and low (7.5 L/min, 30 mmHg). Two resistance clamps (sudden vs. gradual constriction) were compared at the medium resistance setting. Hemolysis was quantified by measuring plasma free hemoglobin (pfHb) and normalized index of hemolysis (NIH). Furthermore, computational fluid dynamics simulations were used to predict the hemolysis metric for each resistance setting with 3D-recreated replicas. Compared to hemolysis metrics reported for state-of-the-art blood pumps, resistance clamps may affect the hemolysis substantially, with up to 84% of the NIH at medium resistance settings and < 47% at low/high resistance settings, regardless of the resistance shape/profile. Numerically predicted hemolysis metric strongly correlated with experimental results for three resistance settings (r > 0.998). The findings underscore the importance of considering the influence of adjustable resistance clamps while investigating the hemolytic potential of blood pumps.

Scientific ReportsVol. 15(1)
Christian Doppler Laboratory for Thermoelectricity (AT), Medical University of Vienna (AT)
Österreichische Nationalstiftung für Forschung, Technologie und Entwicklung, Christian Doppler Forschungsgesellschaft
Openalex Percentile: Top 24%
Mechanical Circulatory Support Devices
0.67
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.